Oxford Nanopore Support | Oxford Nanopore Technologies
Contact support
You can contact our Customer Services and Technical Services teams round the clock using the contact support function.
My PromethION flow cell had fewer than 5000 pores after the flow cell check, what can I do?
My PromethION flow cell had fewer than 5000 pores after the flow cell check, what can I do?
If a PromethION Flow Cell falls below the 5000-pore warranty level, please move it to a different position on the sequencer and re-try the QC check. If it passes, please proceed with sequencing.
If 0 pore was reported, this could be caused by the presence of a bubble between the electrode and the flow cell array.
In this case, flushing through 200 µL of Storage Buffer from the Wash Kit may help.
Please follow the below steps then repeat the platform QC to see if you obtain different results.
Turn the valve to expose the inlet port
After opening the valve, a small tract of air will be visible beyond the inlet port. Drawback a small volume to remove any air bubbles:
Set a P1000 pipette to 200 µL
Insert the tip into the inlet port
Turn the wheel until the dial shows 220-230 µL, or until you can see a small volume of buffer entering the pipette tip
Using a P1000 pipette load 200 µL of Storage Buffer into the inlet port of the flow cell, avoiding the introduction of air bubbles
Close the valve
If the pore count remains below the warranty level after performing the above steps, please contact the support team via Live Chat on our website.