Ligation sequencing DNA V14 - dual barcoding (SQK-NBD114.24 with EXP-PBC096)

概要

  • For barcoding genomic or amplicon DNA for nanopore sequencing.
  • Offering highest accuracy
  • Multiplexing up to 2,304 samples
  • Requires PCR steps
  • Compatible with R10.4.1 flow cells

For Research Use Only

Document version: DBC_9184_v114_revK_20Nov2024

1. Overview of the protocol

Dual Barcoding V14 features:

This protocol requires the use of two kits to generate the dual barcoded libraries:

  • PCR Barcoding Expansion 1-96 (EXP-PBC096): up to 96 unique barcodes are available. These PCR barcodes are used as the primary "inner" barcodes.
  • Native Barcoding Kit 24 V14 (SQK-NBD114.24): up to 24 unique barcodes are available. These native barcodes are used as the secondary "outer" barcodes.

These kits are used in successive order and recommended for users who:
  • Want to multiplex up to 2,304 samples, depending on the barcodes they are using from each expansion.
  • Would like to achieve raw read sequencing modal accuracy of Q20+ (99%) or above.
  • Require control over read length.
  • Would like to utilise upstream processes such as size selection or whole genome amplification.

Introduction to the V14 dual barcoding protocol

This protocol allows massively parallel sequencing of up to 2,304 samples (gDNA or amplicons) on a single flow cell. It uses both the PCR Barcoding Expansion 1-96 (EXP-PBC096) and the Native Barcoding Kit 24 V14 (SQK-NBD114.24).

Samples are initially PCR barcoded with the PCR Barcoding Expansion 1-96 (EXP-PBC096), allowing up to 96 samples to be pooled together. There can be up to 24 pools of 96 samples. Each pool of 96 samples will then undergo secondary barcoding through the ligation of one of 24 native barcodes using the Native Barcoding Kit 24 V14 (SQK-NBD114.24). After secondary barcoding, all pools are combined into a single library for sequencing.

Note: For amplicon inputs, first-round PCR product with the following tailed primers are required. 5’ TTTCTGTTGGTGCTGATATTGC-[ project-specific forward primer sequence ] 3’ 5’ ACTTGCCTGTCGCTCTATCTTC-[ project-specific reverse primer sequence ] 3’

Steps in the sequencing workflow:

Prepare for your experiment

You will need to:

  • Extract your DNA, and check its length, quantity and purity. The quality checks performed during the protocol are essential in ensuring experimental success.
  • Ensure you have your sequencing kit, the correct equipment and third-party reagents
  • Download the software for acquiring and analysing your data
  • Check your flow cell to ensure it has enough pores for a good sequencing run

Library preparation

You will need to:

  • Prepare the DNA ends for adapter attachment
  • Attach barcoding adapters supplied in the 96 PCR Barcoding kit to the DNA ends
  • Amplify each barcoded sample by PCR, then pool the 96 samples together (up to 24 pools of 96 samples can be generated)
  • Prepare the DNA ends of your pooled samples for native barcode attachment
  • Ligate native barcodes to the DNA ends for each pool of 96 samples
  • Pool up to 24 native barcoded libraries together
  • Attach sequencing adapters supplied in the kit to the DNA ends of your combined dual barcoded libraries
  • Prime the flow cell, and load your dual barcoded DNA library into the flow cell Dual barcoding workflow V14 SD edit realigned LH

Sequencing and analysis

You will need to:

  • In the current MinKNOW software version, we recommend setting up live basecalling during the sequencing run without live barcoding. Demultiplexing of the dual barcoded reads will be carried out post-run:
    • Start a sequencing run in the MinKNOW software using SQK-LSK114, which will collect raw data from the device and basecall the reads.
    • Demultiplex your run post-sequencing unsing the MinKNOW software.
  • Start the EPI2ME software and select the barcoding workflow for further analysis (this step is optional).
重要

We do not recommend mixing barcoded libraries with non-barcoded libraries prior to sequencing.

重要

Compatibility of this protocol

This protocol should only be used in combination with:

2. Equipment and consumables

材料
  • 100 ng of each sheared DNA sample to be barcoded in 45 µl
  • OR 100 ng first-round PCR product (with tailed primers) per sample
  • PCR Barcoding Expansion 1-96 (EXP-PBC096)
  • Native Barcoding Kit 24 V14 (SQK-NBD114.24)

消耗品
  • MinION and GridION Flow Cell
  • NEBNext Ultra II End repair/dA-tailing Module (NEB, E7546)
  • NEBNext® Quick Ligation Module (NEB, E6056)
  • LongAmp Hot Start Taq 2X Master Mix (NEB, M0533)
  • NEB Blunt/TA Ligase Master Mix (NEB, M0367)
  • 1.5 ml Eppendorf DNA LoBind tubes
  • 0.2 ml thin-walled PCR tubes or 0.2 ml 96-well PCR plate
  • Freshly prepared 80% ethanol in nuclease-free water
  • Agencourt AMPure XP beads (Beckman Coulter, A63881)
  • Nuclease-free water (e.g. ThermoFisher, AM9937)
  • Qubit™ Assay Tubes (Invitrogen, Q32856)
  • Qubit dsDNA HS Assay Kit (Invitrogen, Q32851)
  • Bovine Serum Albumin (BSA) (50 mg/ml) (e.g Invitrogen™ UltraPure™ BSA 50 mg/ml, AM2616)

装置
  • MinION or GridION device
  • MinION and GridION Flow Cell Light Shield
  • Hula mixer (gentle rotator mixer)
  • Microfuge
  • Microplate centrifuge, e.g. Fisherbrand™ Mini Plate Spinner Centrifuge (Fisher Scientific, 11766427)
  • Vortex mixer
  • Thermal cycler
  • Magnetic rack
  • Multichannel pipette and tips
  • P1000 pipette and tips
  • P200 pipette and tips
  • P100 pipette and tips
  • P20 pipette and tips
  • P10 pipette and tips
  • P2 pipette and tips
  • Ice bucket with ice
  • Timer
  • Qubit fluorometer (or equivalent for QC check)
オプション装置
  • Agilent Bioanalyzer (or equivalent)
  • Eppendorf 5424 centrifuge (or equivalent)

100 ng of gDNA is required per sample.

If using amplicons, 100 ng of first-round PCR product (with tailed primers) is required per sample.

重要

Fragmentation and size selection

For successful amplification, it is critical that the DNA fragment distribution of templates used in the PCR are <8 Kbp. If the input DNA is not < Kbp, use follow steps outline in Shearing genomic DNA using the Covaris g-TUBE™ to shear the sample to each a fragment distribution <8 Kbp.

Additionally, we offer several options for size-selecting your DNA sample to enrich for long fragments. Instructions are available in the Size Selection section of Extraction methods.

Input DNA

How to QC your input DNA

It is important that the input DNA meets the quantity and quality requirements. Using too little or too much DNA, or DNA of poor quality (e.g. highly fragmented or containing RNA or chemical contaminants) can affect your library preparation.

For instructions on how to perform quality control of your DNA sample, please read the Input DNA/RNA QC protocol.

Chemical contaminants

Depending on how the DNA is extracted from the raw sample, certain chemical contaminants may remain in the purified DNA, which can affect library preparation efficiency and sequencing quality. Read more about contaminants on the Contaminants page of the Community.

Third-party reagents

We have validated and recommend the use of all the third-party reagents used in this protocol. Alternatives have not been tested by Oxford Nanopore Technologies.

For all third-party reagents, we recommend following the manufacturer's instructions to prepare the reagents for use.

Check your flow cell

We highly recommend that you check the number of pores in your flow cell prior to starting a sequencing experiment. This should be done within 12 weeks of purchasing for MinION/GridION/PromethION or within four weeks of purchasing Flongle Flow Cells. Oxford Nanopore Technologies will replace any flow cell with fewer than the number of pores in the table below, when the result is reported within two days of performing the flow cell check, and when the storage recommendations have been followed. To do the flow cell check, please follow the instructions in the Flow Cell Check document.

Flow cell Minimum number of active pores covered by warranty
Flongle Flow Cell 50
MinION/GridION Flow Cell 800
PromethION Flow Cell 5000
重要

AMPure XP Beads

Within the Native Barcoding Kit 24 V14 (SQK-NBD114.24), AMPure XP Beads (AXP) are supplied at the volume needed to complete the native barcoding sections of the protocol: the second "End-prep", "Native barcode ligation" and "Adapter ligation and clean-up".

However, extra AMPure XP Beads are required for the PCR barcoding steps of the protocol: the first "End-prep", "Ligation of Barcode Adapter" and "Barcoding PCR".

Please note, other purification methods are available.

PCR Barcoding Expansion Pack 1-96 (EXP-PBC096)

2655 10999
Name Acronym Cap colour No. of vials/plates Fill volume per well (µl)
PCR Barcode Primer Mix plate BC01-96 White 1 plate 24
Barcode Adapter plate BCA Blue 1 plate 240

Layout of barcodes in the 96 tube plate

The wells of the 96 tube plate correspond to the barcodes in the following way. All barcodes are supplied at 10 µM concentration and to be used at a final concentration of 0.2 µM.

Barcode 96 plate

Capping and decapping the 96 well format

重要

The Native Adapter (NA) used in this kit and protocol is not interchangeable with other sequencing adapters.

Native Barcoding Kit 24 V14 (SQK-NBD114.24) contents

Note: We are in the process of reformatting the barcodes provided in this kit into a plate format. This will reduce plastic waste and will facilitate automated applications.

Plate format

SQK-NBD114.24 plate format

Name Acronym Cap colour No. of vials Fill volume per vial (µl)
DNA Control Sample DCS Yellow 2 35
Native Adapter NA Green 1 40
Sequencing Buffer SB Red 1 700
Library Beads LIB Pink 1 600
Library Solution LIS White cap, pink label 1 600
Elution Buffer EB Black 2 500
AMPure XP Beads AXP Clear cap, light teal label 1 6,000
Long Fragment Buffer LFB Orange 1 1,800
Short Fragment Buffer SFB Clear 1 1,800
EDTA EDTA Blue 1 700
Flow Cell Flush FCF Clear cap, light blue label 1 8,000
Flow Cell Tether FCT Purple 1 200
Native Barcode plate NB01-24 - 2 plates, 3 sets of barcodes per plate 5 µl per well

Note: This Product Contains AMPure XP Reagent Manufactured by Beckman Coulter, Inc. and can be stored at -20°C with the kit without detriment to reagent stability.

Note: The DNA Control Sample (DCS) is a 3.6 kb standard amplicon mapping the 3' end of the Lambda genome.


Vial format

SQK-NBD114.24 bottle format

Name Acronym Cap colour No. of vials Fill volume per vial (µl)
Native Barcodes NB01-24 Clear 24 (one per barcode) 20
DNA Control Sample DCS Yellow 2 35
Native Adapter NA Green 1 40
Sequencing Buffer SB Red 1 700
Library Beads LIB Pink 1 600
Library Solution LIS White cap, pink label 1 600
Elution Buffer EB Black 2 500
AMPure XP Beads AXP Clear cap, light teal label 1 6,000
Long Fragment Buffer LFB Orange 1 1,800
Short Fragment Buffer SFB Clear 1 1,800
EDTA EDTA Blue 1 700
Flow Cell Flush FCF Clear cap, light blue label 1 8,000
Flow Cell Tether FCT Purple 1 200

Note: This Product Contains AMPure XP Reagent Manufactured by Beckman Coulter, Inc. and can be stored at -20°C with the kit without detriment to reagent stability.

Note: The DNA Control Sample (DCS) is a 3.6 kb standard amplicon mapping the 3' end of the Lambda genome.

To maximise the use of the Native Barcoding Kits, the Native Barcode Auxiliary V14 (EXP-NBA114) and the Sequencing Auxiliary Vials V14 (EXP-AUX003) expansion packs are available.

These expansions provide extra library preparation and flow cell priming reagents to allow users to utilise any unused barcodes for those running in smaller subsets.

Both expansion packs used together will provide enough reagents for 12 reactions. For customers requiring extra EDTA to maximise the use of barcodes, we recommend using 0.25 M EDTA and adding 4 µl for library preps using the SQK-NBD114.24 kit.

Native Barcode Auxiliary V14 (EXP-NBA114) contents:

EXP-NBA114 tubes

Note: This Product contains AMPure XP Reagent manufactured by Beckman Coulter, Inc. and can be stored at -20°C with the kit without detriment to reagent stability.

Sequencing Auxiliary Vials V14 (EXP-AUX003) contents:

EXP-AUX003 bottles

96 barcode sequences

Component Sequence
BC01 / RB01 AAGAAAGTTGTCGGTGTCTTTGTG
BC02 / RB02 TCGATTCCGTTTGTAGTCGTCTGT
BC03 / RB03 GAGTCTTGTGTCCCAGTTACCAGG
BC04 / RB04 TTCGGATTCTATCGTGTTTCCCTA
BC05 / RB05 CTTGTCCAGGGTTTGTGTAACCTT
BC06 / RB06 TTCTCGCAAAGGCAGAAAGTAGTC
BC07 / RB07 GTGTTACCGTGGGAATGAATCCTT
BC08 / RB08 TTCAGGGAACAAACCAAGTTACGT
BC09 / RB09 AACTAGGCACAGCGAGTCTTGGTT
BC10 / RB10 AAGCGTTGAAACCTTTGTCCTCTC
BC11 / RB11 GTTTCATCTATCGGAGGGAATGGA
BC12 / RB12 CAGGTAGAAAGAAGCAGAATCGGA
BC13 / 16S13 / RB13 AGAACGACTTCCATACTCGTGTGA
BC14 / 16S14 / RB14 AACGAGTCTCTTGGGACCCATAGA
BC15 / 16S15 / RB15 AGGTCTACCTCGCTAACACCACTG
BC16 / 16S16 / RB16 CGTCAACTGACAGTGGTTCGTACT
BC17 / 16S17 / RB17 ACCCTCCAGGAAAGTACCTCTGAT
BC18 / 16S18 / RB18 CCAAACCCAACAACCTAGATAGGC
BC19 / 16S19 / RB19 GTTCCTCGTGCAGTGTCAAGAGAT
BC20 / 16S20 / RB20 TTGCGTCCTGTTACGAGAACTCAT
BC21 / 16S21 / RB21 GAGCCTCTCATTGTCCGTTCTCTA
BC22 / 16S22 / RB22 ACCACTGCCATGTATCAAAGTACG
BC23 / 16S23 / RB23 CTTACTACCCAGTGAACCTCCTCG
BC24 / 16S24 / RB24 GCATAGTTCTGCATGATGGGTTAG
BC25 / RB25 GTAAGTTGGGTATGCAACGCAATG
BC26 / RB26 CATACAGCGACTACGCATTCTCAT
BC27 / RB27 CGACGGTTAGATTCACCTCTTACA
BC28 / RB28 TGAAACCTAAGAAGGCACCGTATC
BC29 / RB29 CTAGACACCTTGGGTTGACAGACC
BC30 / RB30 TCAGTGAGGATCTACTTCGACCCA
BC31 / RB31 TGCGTACAGCAATCAGTTACATTG
BC32 / RB32 CCAGTAGAAGTCCGACAACGTCAT
BC33 / RB33 CAGACTTGGTACGGTTGGGTAACT
BC34 / RB34 GGACGAAGAACTCAAGTCAAAGGC
BC35 / RB35 CTACTTACGAAGCTGAGGGACTGC
BC36 / RB36 ATGTCCCAGTTAGAGGAGGAAACA
BC37 / RB37 GCTTGCGATTGATGCTTAGTATCA
BC38 / RB38 ACCACAGGAGGACGATACAGAGAA
BC39 / RB39 CCACAGTGTCAACTAGAGCCTCTC
BC40 / RB40 TAGTTTGGATGACCAAGGATAGCC
BC41 / RB41 GGAGTTCGTCCAGAGAAGTACACG
BC42 / RB42 CTACGTGTAAGGCATACCTGCCAG
BC43 / RB43 CTTTCGTTGTTGACTCGACGGTAG
BC44 / RB44 AGTAGAAAGGGTTCCTTCCCACTC
BC45 / RB45 GATCCAACAGAGATGCCTTCAGTG
BC46 / RB46 GCTGTGTTCCACTTCATTCTCCTG
BC47 / RB47 GTGCAACTTTCCCACAGGTAGTTC
BC48 / RB48 CATCTGGAACGTGGTACACCTGTA
BC49 / RB49 ACTGGTGCAGCTTTGAACATCTAG
BC50 / RB50 ATGGACTTTGGTAACTTCCTGCGT
BC51 / RB51 GTTGAATGAGCCTACTGGGTCCTC
BC52 / RB52 TGAGAGACAAGATTGTTCGTGGAC
BC53 / RB53 AGATTCAGACCGTCTCATGCAAAG
BC54 / RB54 CAAGAGCTTTGACTAAGGAGCATG
BC55 / RB55 TGGAAGATGAGACCCTGATCTACG
BC56 / RB56 TCACTACTCAACAGGTGGCATGAA
BC57 / RB57 GCTAGGTCAATCTCCTTCGGAAGT
BC58 / RB58 CAGGTTACTCCTCCGTGAGTCTGA
BC59 / RB59 TCAATCAAGAAGGGAAAGCAAGGT
BC60 / RB60 CATGTTCAACCAAGGCTTCTATGG
BC61 / RB61 AGAGGGTACTATGTGCCTCAGCAC
BC62 / RB62 CACCCACACTTACTTCAGGACGTA
BC63 / RB63 TTCTGAAGTTCCTGGGTCTTGAAC
BC64 / RB64 GACAGACACCGTTCATCGACTTTC
BC65 / RB65 TTCTCAGTCTTCCTCCAGACAAGG
BC66 / RB66 CCGATCCTTGTGGCTTCTAACTTC
BC67 / RB67 GTTTGTCATACTCGTGTGCTCACC
BC68 / RB68 GAATCTAAGCAAACACGAAGGTGG
BC69 / RB69 TACAGTCCGAGCCTCATGTGATCT
BC70 / RB70 ACCGAGATCCTACGAATGGAGTGT
BC71 / RB71 CCTGGGAGCATCAGGTAGTAACAG
BC72 / RB72 TAGCTGACTGTCTTCCATACCGAC
BC73 / RB73 AAGAAACAGGATGACAGAACCCTC
BC74 / RB74 TACAAGCATCCCAACACTTCCACT
BC75 / RB75 GACCATTGTGATGAACCCTGTTGT
BC76 / RB76 ATGCTTGTTACATCAACCCTGGAC
BC77 / RB77 CGACCTGTTTCTCAGGGATACAAC
BC78 / RB78 AACAACCGAACCTTTGAATCAGAA
BC79 / RB79 TCTCGGAGATAGTTCTCACTGCTG
BC80 / RB80 CGGATGAACATAGGATAGCGATTC
BC81 / RB81 CCTCATCTTGTGAAGTTGTTTCGG
BC82 / RB82 ACGGTATGTCGAGTTCCAGGACTA
BC83 / RB83 TGGCTTGATCTAGGTAAGGTCGAA
BC84 / RB84 GTAGTGGACCTAGAACCTGTGCCA
BC85 / RB85 AACGGAGGAGTTAGTTGGATGATC
BC86 / RB86 AGGTGATCCCAACAAGCGTAAGTA
BC87 / RB87 TACATGCTCCTGTTGTTAGGGAGG
BC88 / RB88 TCTTCTACTACCGATCCGAAGCAG
BC89 / RB89 ACAGCATCAATGTTTGGCTAGTTG
BC90 / RB90 GATGTAGAGGGTACGGTTTGAGGC
BC91 / RB91 GGCTCCATAGGAACTCACGCTACT
BC92 / RB92 TTGTGAGTGGAAAGATACAGGACC
BC93 / RB93 AGTTTCCATCACTTCAGACTTGGG
BC94 / RB94 GATTGTCCTCAAACTGCCACCTAC
BC95 / RB95 CCTGTCTGGAAGAAGAATGGACTT
BC96 / RB96 CTGAACGGTCATAGAGTCCACCAT

Native barcode sequences

Component Forward sequence Reverse sequence
NB01 CACAAAGACACCGACAACTTTCTT AAGAAAGTTGTCGGTGTCTTTGTG
NB02 ACAGACGACTACAAACGGAATCGA TCGATTCCGTTTGTAGTCGTCTGT
NB03 CCTGGTAACTGGGACACAAGACTC GAGTCTTGTGTCCCAGTTACCAGG
NB04 TAGGGAAACACGATAGAATCCGAA TTCGGATTCTATCGTGTTTCCCTA
NB05 AAGGTTACACAAACCCTGGACAAG CTTGTCCAGGGTTTGTGTAACCTT
NB06 GACTACTTTCTGCCTTTGCGAGAA TTCTCGCAAAGGCAGAAAGTAGTC
NB07 AAGGATTCATTCCCACGGTAACAC GTGTTACCGTGGGAATGAATCCTT
NB08 ACGTAACTTGGTTTGTTCCCTGAA TTCAGGGAACAAACCAAGTTACGT
NB09 AACCAAGACTCGCTGTGCCTAGTT AACTAGGCACAGCGAGTCTTGGTT
NB10 GAGAGGACAAAGGTTTCAACGCTT AAGCGTTGAAACCTTTGTCCTCTC
NB11 TCCATTCCCTCCGATAGATGAAAC GTTTCATCTATCGGAGGGAATGGA
NB12 TCCGATTCTGCTTCTTTCTACCTG CAGGTAGAAAGAAGCAGAATCGGA
NB13 AGAACGACTTCCATACTCGTGTGA TCACACGAGTATGGAAGTCGTTCT
NB14 AACGAGTCTCTTGGGACCCATAGA TCTATGGGTCCCAAGAGACTCGTT
NB15 AGGTCTACCTCGCTAACACCACTG CAGTGGTGTTAGCGAGGTAGACCT
NB16 CGTCAACTGACAGTGGTTCGTACT AGTACGAACCACTGTCAGTTGACG
NB17 ACCCTCCAGGAAAGTACCTCTGAT ATCAGAGGTACTTTCCTGGAGGGT
NB18 CCAAACCCAACAACCTAGATAGGC GCCTATCTAGGTTGTTGGGTTTGG
NB19 GTTCCTCGTGCAGTGTCAAGAGAT ATCTCTTGACACTGCACGAGGAAC
NB20 TTGCGTCCTGTTACGAGAACTCAT ATGAGTTCTCGTAACAGGACGCAA
NB21 GAGCCTCTCATTGTCCGTTCTCTA TAGAGAACGGACAATGAGAGGCTC
NB22 ACCACTGCCATGTATCAAAGTACG CGTACTTTGATACATGGCAGTGGT
NB23 CTTACTACCCAGTGAACCTCCTCG CGAGGAGGTTCACTGGGTAGTAAG
NB24 GCATAGTTCTGCATGATGGGTTAG CTAACCCATCATGCAGAACTATGC

3. End-prep

材料
  • 100 ng of each sheared DNA sample to be barcoded in 45 µl

消耗品
  • NEBNext® Ultra II End Prep Enzyme Mix from NEBNext® Ultra II End Repair Module (NEB, E7546)
  • NEBNext® Ultra II End Prep Reaction Buffer from NEBNext® Ultra II End Repair Module (NEB, E7546)
  • Freshly prepared 80% ethanol in nuclease-free water
  • Nuclease-free water (e.g. ThermoFisher, AM9937)
  • Agencourt AMPure XP beads (Beckman Coulter, A63881)
  • 0.2 ml thin-walled PCR tubes or 0.2 ml 96-well PCR plate
  • Qubit™ Assay Tubes (Invitrogen, Q32856)
  • Qubit dsDNA HS Assay Kit (Invitrogen, Q32851)

装置
  • P1000 pipette and tips
  • P100 pipette and tips
  • P10 pipette and tips
  • Thermal cycler
  • Ice bucket with ice
  • Microfuge
  • Magnetic rack
  • Vortex mixer
オプション装置
  • Qubit fluorometer (or equivalent for QC check)
CHECKPOINT

Check your flow cell.

We recommend performing a flow cell check before starting your library prep to ensure you have a flow cell with enough pores for a good sequencing run.

See the flow cell check instructions in the MinKNOW protocol for more information.

Prepare the NEBNext Ultra II End Repair / dA-tailing Module reagents in accordance with manufacturer's instructions, and place on ice:

For optimal performance, NEB recommend the following:

  1. Thaw all reagents on ice.

  2. Ensure the reagents are well mixed.
    Note: Do not vortex the Ultra II End Prep Enzyme Mix.

  3. Always spin down tubes before opening for the first time each day.

  4. The NEBNext Ultra II End Prep Reaction Buffer may contain a white precipitate. If this occurs, allow the mixture(s) to come to room temperature and pipette the buffer several times to break up the precipitate, followed by a quick vortex to mix.

Prepare the DNA in nuclease-free water.

  1. Transfer 100 ng DNA of each sample into a fresh 0.2 ml PCR tube or plate
  2. Adjust the volume to 45 μl with nuclease-free water
  3. Mix thoroughly by flicking the tube to avoid unwanted shearing
  4. Spin down briefly in a microfuge

Set up the end-repair reaction as follows for each library:

Reagent Volume per sample
100 ng DNA 45 µl
Ultra II End-prep reaction buffer 7 µl
Ultra II End-prep enzyme mix 3 µl
Nuclease-free water 5 µl
Total 60 µl

Mix by pipetting and briefly spin down.

Using a thermal cycler, incubate for 5 minutes at 20 °C and 5 minutes at 65 °C.

Resuspend the AMPure XP beads by vortexing.

Add 60 µl of resuspended AMPure XP beads to the end-prep reaction and mix by pipetting.

Incubate at room temperature for 5 minutes.

Prepare sufficient fresh 80% ethanol in nuclease-free water for all of your samples. Allow enough for 400 µl per sample, with some excess.

Spin down the samples and pellet on a magnet until supernatant is clear and colourless. Keep the samples on the magnet, and pipette off the supernatant.

Keep the samples on the magnet and wash the beads with 200 µl of freshly prepared 80% ethanol without disturbing the pellet. Remove the ethanol using a pipette and discard.

Repeat the previous step.

Spin down and place the samples back on the magnetic rack. Pipette off any residual ethanol. Allow the pellet to dry for ~30 seconds, but do not dry the pellet to the point of cracking.

Remove the samples from the magnet and resuspend each pellet in 16 µl nuclease-free water. Incubate for 2 minutes at room temperature.

Pellet the beads on a magnet until the eluate is clear and colourless.

Remove eluate once it is clear and colourless. Transfer each eluted sample to a new tube or plate well.

Quantify 1 µl of end-prepped DNA using a Qubit fluorometer.

最終ステップ

Take forward the end-prepped DNA into the next step. However, at this point it is also possible to store the sample at 4°C overnight.

4. Ligation of Barcode Adapter

材料
  • Barcode Adapter (BCA)

消耗品
  • NEB Blunt/TA Ligase Master Mix (NEB, cat # M0367)
  • Agencourt AMPure XP beads (Beckman Coulter, A63881)
  • Nuclease-free water (e.g. ThermoFisher, AM9937)
  • Freshly prepared 80% ethanol in nuclease-free water
  • 0.2 ml thin-walled PCR tubes or 0.2 ml 96-well PCR plate
  • Qubit™ Assay Tubes (Invitrogen, Q32856)
  • Qubit dsDNA HS Assay Kit (Invitrogen, Q32851)

装置
  • Microfuge
  • Hula mixer (gentle rotator mixer)
  • Vortex mixer
  • Ice bucket with ice
  • Multichannel pipette and tips
  • Magnetic rack
  • P1000 pipette and tips
  • P100 pipette and tips
  • P10 pipette and tips
  • Qubit fluorometer (or equivalent for QC check)

Prepare the NEB Blunt/TA Ligase Master Mix according to the manufacturer's instructions, and place on ice:

  1. Thaw the reagents at room temperature.

  2. Spin down the reagent tubes for 5 seconds.

  3. Ensure the reagents are fully mixed by performing 10 full volume pipette mixes.

Spin down the Barcode Adapter (BCA), pipette mix and place on ice.

Add the reagents in the order given below, into fresh 0.2 ml PCR tubes or 96-well plate:

Reagent Volume
End-prepped DNA 15 µl
Barcode Adapter 10 µl
Blunt/TA Ligase Master Mix 25 µl
Total 50 µl

Mix by pipetting and briefly spin down.

Incubate the samples for 10 minutes at room temperature.

Resuspend the AMPure XP beads by vortexing.

Add 20 µl of resuspended AMPure XP beads to each sample for a 0.4X clean and mix by pipetting up and down ten times.

Incubate on a Hula mixer (rotator mixer) for 5 minutes at room temperature.

Prepare sufficient fresh 80% ethanol in nuclease-free water for all of your samples. Allow enough for 400 µl per sample, with some excess.

Place on a magnetic rack, allow beads to pellet and pipette off supernatant.

Keep the samples on the magnet and wash the beads with 200 µl of freshly prepared 80% ethanol without disturbing the pellet. Remove the ethanol using a pipette and discard.

Repeat the previous step.

Place the samples back on the magnet. Pipette off any residual 80% ethanol. Allow to dry for ~30 seconds, but do not dry the pellet to the point of cracking.

Remove the samples from the magnet and resuspend pellet in 25 µl nuclease-free water. Incubate for 2 minutes at room temperature.

Pellet the beads on a magnet until the eluate is clear and colourless.

Remove and retain the eluate once it is clear and colourless. Transfer each eluted sample to a fresh 0.2 ml PCR tube or plate.

  • Dispose of the pelleted beads.

Quantify 1 µl of the adapter ligated DNA using a Qubit fluorometer.

最終ステップ

Take forward the adapter ligated samples into the Barcoding PCR step. However, at this point it is also possible to store the sample at 4°C overnight.

5. Barcoding PCR

材料
  • PCR Barcodes (BC01-96, at 10 µM)

消耗品
  • LongAmp Hot Start Taq 2X Master Mix (NEB, M0533)
  • Nuclease-free water (e.g. ThermoFisher, AM9937)
  • 1.5 ml Eppendorf DNA LoBind tubes
  • Eppendorf twin.tec® PCR plate 96 LoBind, semi-skirted (Eppendorf™, cat # 0030129504) with heat seals
  • OR 0.2 ml thin-walled PCR tubes
  • Agencourt AMPure XP Beads (Beckman Coulter™, A63881)
  • Freshly prepared 80% ethanol in nuclease-free water
  • Qubit™ Assay Tubes (Invitrogen, Q32856)
  • Qubit dsDNA HS Assay Kit (Invitrogen, Q32851)

装置
  • Thermal cycler
  • Magnetic rack
  • Microfuge
  • P1000 pipette and tips
  • P100 pipette and tips
  • P10 pipette and tips
  • Qubit fluorometer (or equivalent for QC check)

Please note, this protocol is written for a template input of 100–200 fmol with PCR Barcodes (BC01-96) used at a final concentration of 0.2 µM. However, the input mass and the number of PCR cycles may be adjusted as appropriate depending on the requirements of the experiment.

Thaw the PCR Barcodes (BC01-96) required for your number of samples at room temperature. Individually mix the barcodes by pipetting, spin down, and place on ice.

重要

If using amplicon samples, ensure the samples have undergone a round of PCR with tailed primers before commencing with the protocol.

Prepare the samples in nuclease-free water:

  1. Transfer 100-200 fmol of each sample to a clear 0.2 ml PCR tube or plate
  • For 1–12 samples: Adjust the volume to 48 μl with nuclease-free water
  • For 13–96 samples: Adjust the volume to 24 μl with nuclease-free water
  1. Mix thoroughly by flicking the tube or plate to avoid unwanted shearing
  2. Spin down briefly in a microfuge

Select a unique barcode for each sample to be processed in the PCR barcoded pool.

Note: Only use one barcode per sample.

Set up a barcoding PCR reaction as follows for each library in fresh 0.2 ml PCR tubes or a 0.2 ml 96-well PCR plate.

Between each addition, pipette mix 10-20 times.

Reagent Volume per sample for using 1–12 barcodes Volume per sample for using 13 barcodes or more
PCR Barcode (one of BC1-BC96, at 10 µM) 2 µl 1 µl
Adapter-ligated DNA 48 µl 24 µl
LongAmp Taq 2x master mix 50 µl 25 µl
Total volume 100 µl 50 µl

Mix by pipetting and briefly spin down.

Amplify using the following cycling conditions:

Cycle step Temperature Time No. of cycles
Initial denaturation 94 °C 3 mins 1
Denaturation 94 °C 15 secs 15-18 (a)
Annealing 56 °C 15 secs 15-18 (a)
Extension 65 °C 6 mins (b) 15-18 (a)
Final extension 65 °C 10 mins 1
Hold 4 °C

a. Adjust accordingly if input quantities are altered.

b. Adjust accordingly for different lengths of amplicons and the type of polymerase that is being used (LongAmp Taq amplifies at a rate of 50 seconds per kb). Here 6 min is used for ~8 Kbp templates.

Resuspend the AMPure XP beads by vortexing.

Add 0.4X volume of resuspended AMPure XP Beads to each reaction and mix by flicking the tube.

Reagent Volume for 100 µl samples Volume for 50 µl samples
AMPure XP Beads 40 µl 20 µl

Incubate at room temperature for 5 minutes.

Prepare sufficient fresh 80% ethanol in nuclease-free water for all of your samples. Allow enough for 400 µl per sample, with some excess.

Place samples on a magnetic rack, allow beads to pellet and pipette off supernatant.

Keep the samples on the magnet and wash the beads with 200 µl of freshly prepared 80% ethanol without disturbing the pellets. Remove the ethanol using a pipette and discard.

Repeat the previous step.

Spin down and place the samples back on the magnet. Pipette off any residual ethanol. Allow to dry for ~30 seconds, but do not dry the pellets to the point of cracking.

Remove the samples from the magnetic rack and resuspend each pellet in 10 µl nuclease-free water. Incubate for 2 minutes at room temperature.

Pellet the beads on a magnetic rack until the eluate is clear and colourless.

Remove and retain 10 µl of each eluate into clean 0.2 ml PCR tubes or a clean PCR plate.

  • Dispose of the pelleted beads

Quantify the PCR barcoded samples using a Qubit fluorometer and pool all barcoded samples in the desired ratios into a 1.5 ml DNA LoBind Eppendorf tube for each PCR barcoded sample pool.

重要

Each PCR barcoded sample pool should contain 1–96 unique barcodes.

You will have up to 24 separate pools of 96 samples going into the end-prep and native barcode ligation section of the protocol.

Note: Do not mix different PCR barcoded sample pools prior to secondary "outer" barcoding.

Prepare each pooled PCR barcoded sample pool to 200 fmol (130 ng for 1 kb amplicons) in separate tubes and make the volume up to 12.5 µl nuclease-free water.

If the volume of a pool exceeds the 12.5 µl required for the end-prep reaction, consider a 2.5X AMPure XP Bead purification of the pool to concentrate your sample.

最終ステップ

The pooled barcoded libraries are now ready to be end-prepped and undergo secondary barcoding. However, at this point it is also possible to store the libraries at 4°C overnight.

6. End-prep

材料
  • Up to 24 pools of PCR barcoded sample pools (with up to 96 samples in each pool), each in 12.5 µl
  • AMPure XP Beads (AXP)

消耗品
  • NEBNext® Ultra II End Prep Enzyme Mix from NEBNext® Ultra II End Repair Module (NEB, E7546)
  • NEBNext® Ultra II End Prep Reaction Buffer from NEBNext® Ultra II End Repair Module (NEB, E7546)
  • Freshly prepared 80% ethanol in nuclease-free water
  • 1.5 ml Eppendorf DNA LoBind tubes
  • Eppendorf twin.tec® PCR plate 96 LoBind, semi-skirted (Eppendorf™, cat # 0030129504) with heat seals
  • OR 0.2 ml thin-walled PCR tubes
  • Nuclease-free water (e.g. ThermoFisher, AM9937)
  • Qubit™ Assay Tubes (Invitrogen, Q32856)
  • Qubit dsDNA HS Assay Kit (Invitrogen, Q32851)

装置
  • P1000 pipette and tips
  • P200 pipette and tips
  • P100 pipette and tips
  • P20 pipette and tips
  • P10 pipette and tips
  • P2 pipette and tips
  • Multichannel pipette and tips
  • Thermal cycler
  • Microplate centrifuge, e.g. Fisherbrand™ Mini Plate Spinner Centrifuge (Fisher Scientific, 11766427)
  • Microfuge
  • Ice bucket with ice
  • Magnetic rack
  • Vortex mixer
  • Hula mixer (rotator mixer)
  • Qubit fluorometer (or equivalent)

Thaw the AMPure XP Beads (AXP) at room temperature and mix by vortexing. Keep the beads at room temperature until use.

Prepare the NEBNext Ultra II End Repair / dA-tailing Module reagents in accordance with manufacturer's instructions, and place on ice:

For optimal performance, NEB recommend the following:

  1. Thaw all reagents on ice.

  2. Ensure the reagents are well mixed.
    Note: Do not vortex the Ultra II End Prep Enzyme Mix.

  3. Always spin down tubes before opening for the first time each day.

  4. The NEBNext Ultra II End Prep Reaction Buffer may contain a white precipitate. If this occurs, allow the mixture(s) to come to room temperature and pipette the buffer several times to break up the precipitate, followed by a quick vortex to mix.

In clean 0.2 ml thin-walled PCR tubes (or a clean 96-well plate), prepare 200 fmol (130 ng for 1 kb amplicons) of each PCR barcoded sample pool.

Make up each PCR barcoded sample pool to 12.5 µl using nuclease-free water. Mix gently by pipetting and spin down.

Combine the following components per tube/well:

Between each addition, pipette mix 10 - 20 times.

Reagent Volume
PCR barcoded sample pool 12.5 µl
Ultra II End-prep Reaction Buffer 1.75 µl
Ultra II End-prep Enzyme Mix 0.75 µl
Total 15 µl
ヒント

We recommend making up a mastermix of the end-prep and DNA repair reagents for the total number of PCR barcoded sample pools and adding 2.5 µl to each well.

Ensure the components are thoroughly mixed by pipetting and spin down in a centrifuge.

Using a thermal cycler, incubate at 20°C for 5 minutes and 65°C for 5 minutes.

Transfer each PCR barcoded sample pool into a clean 1.5 ml Eppendorf DNA LoBind tube.

Resuspend the AMPure XP beads (AXP) by vortexing.

Add 15 µl of resuspended AMPure XP Beads (AXP) to each end-prep reaction and mix by flicking the tube.

Incubate on a Hula mixer (rotator mixer) for 5 minutes at room temperature.

Prepare sufficient fresh 80% ethanol in nuclease-free water for all of your samples. Allow enough for 400 µl per PCR barcoded sample pool, with some excess.

Spin down the samples and pellet the beads on a magnet until the eluate is clear and colourless. Keep the tubes on the magnet and pipette off the supernatant.

Keep the tubes on the magnet and wash the beads with 200 µl of freshly prepared 80% ethanol without disturbing the pellet. Remove the ethanol using a pipette and discard.

If the pellet was disturbed, wait for beads to pellet again before removing the ethanol.

Repeat the previous step.

Briefly spin down and place the tubes back on the magnet for the beads to pellet. Pipette off any residual ethanol. Allow to dry for 30 seconds, but do not dry the pellets to the point of cracking.

Remove the tubes from the magnetic rack and resuspend the pellet in 10 µl nuclease-free water. Spin down and incubate for 2 minutes at room temperature.

Pellet the beads on a magnet until the eluate is clear and colourless.

Remove and retain 10 µl of eluate into a clean 1.5 ml Eppendorf DNA LoBind tube.

  • Dispose of the pelleted beads
CHECKPOINT

Quantify 1 µl of each eluted end-prepped PCR barcoded sample pool using a Qubit fluorometer.

重要

You will have up to 24 separate end-prepped PCR barcoded sample pools to take forward into secondary barcoding via native barcode ligation.

Note: Do not mix the PCR barcoded sample pools prior to secondary "outer" barcoding.

最終ステップ

Take forward an equimolar mass of each of the end-prepped PCR barcoded sample pools to undergo secondary barcoding in the native barcode ligation step. However, at this point it is also possible to store the PCR barcoded sample pools at 4°C overnight.

7. Native barcode ligation

材料
  • Native Barcodes (NB01-24)
  • AMPure XP Beads (AXP)
  • EDTA (EDTA)

消耗品
  • NEB Blunt/TA Ligase Master Mix (NEB, M0367)
  • Freshly prepared 80% ethanol in nuclease-free water
  • Nuclease-free water (e.g. ThermoFisher, AM9937)
  • 1.5 ml Eppendorf DNA LoBind tubes
  • Eppendorf twin.tec® PCR plate 96 LoBind, semi-skirted (Eppendorf™, cat # 0030129504) with heat seals
  • OR 0.2 ml thin-walled PCR tubes
  • Qubit™ Assay Tubes (Invitrogen, Q32856)
  • Qubit dsDNA HS Assay Kit (ThermoFisher, cat # Q32851)

装置
  • Magnetic rack
  • Vortex mixer
  • Hula mixer (gentle rotator mixer)
  • Microfuge
  • Thermal cycler
  • Ice bucket with ice
  • Multichannel pipette and tips
  • P1000 pipette and tips
  • P200 pipette and tips
  • P100 pipette and tips
  • P20 pipette and tips
  • P10 pipette and tips
  • P2 pipette and tips
  • Qubit fluorometer (or equivalent for QC check)

Prepare the NEB Blunt/TA Ligase Master Mix according to the manufacturer's instructions, and place on ice:

  1. Thaw the reagents at room temperature.

  2. Spin down the reagent tubes for 5 seconds.

  3. Ensure the reagents are fully mixed by performing 10 full volume pipette mixes.

Thaw the EDTA at room temperature and mix by vortexing. Then spin down and place on ice.

Thaw the Native Barcodes (NB01-24) at room temperature. Briefly spin down, individually mix the barcodes required for your number of PCR barcoded sample pools by pipetting, and place them on ice.

Select a unique barcode for each PCR barcoded sample pool to be run together on the same flow cell. Up to 24 PCR barcoded sample pools can be barcoded and combined in one experiment.

Note: Only use one barcode per PCR barcoded sample pool.

In clean 0.2 ml PCR-tubes or a 96-well plate, add the reagents in the following order per well:

Between each addition, pipette mix 10 - 20 times.

Reagent Volume
End-prepped DNA (PCR barcoded sample pools) 7.5 µl
Native Barcode (NB01-24) 2.5 µl
Blunt/TA Ligase Master Mix 10 µl
Total 20 µl

Thoroughly mix the reaction by gently pipetting and briefly spinning down.

Incubate for 20 minutes at room temperature.

Add the following volume of EDTA to each well and mix thoroughly by pipetting and spin down briefly.

Note: Ensure you follow the instructions for the cap colour of your EDTA tube.

EDTA cap colour Volume per well
For clear cap EDTA 2 µl
For blue cap EDTA 4 µl
ヒント

EDTA is added at this step to stop the reaction.

Pool all the native barcoded sample pools in a 1.5 ml Eppendorf DNA LoBind tube.

Note: Ensure you follow the instructions for the cap colour of your EDTA tube.

Volume per PCR barcoded sample pool For 6 sample pools For 12 sample pools For 24 sample pools
Total volume for preps using clear cap EDTA 22 µl 132 µl 264 µl 528 µl
Total volume for preps using blue cap EDTA 24 µl 144 µl 288 µl 576 µl
ヒント

We recommend checking the base of your tubes/plate are all the same volume before pooling and after to ensure all the liquid has been taken forward.

Resuspend the AMPure XP Beads (AXP) by vortexing.

Add 0.4X AMPure XP Beads (AXP) to the pooled reaction, and mix by pipetting.

Note: Ensure you follow the instructions for the cap colour of your EDTA tube.

Volume per sample For 6 samples For 12 samples For 24 samples
Volume of AXP for preps using clear cap EDTA 9 µl 53 µl 106 µl 211 µl
Volume of AXP for preps using blue cap EDTA 10 µl 58 µl 115 µl 230 µl

Incubate on a Hula mixer (rotator mixer) for 10 minutes at room temperature.

Prepare 2 ml of fresh 80% ethanol in nuclease-free water.

Spin down the sample and pellet on a magnet for 5 minutes. Keep the tube on the magnetic rack until the eluate is clear and colourless, and pipette off the supernatant.

Keep the tube on the magnetic rack and wash the beads with 700 µl of freshly prepared 80% ethanol without disturbing the pellet. Remove the ethanol using a pipette and discard.

If the pellet was disturbed, wait for beads to pellet again before removing the ethanol.

Repeat the previous step.

Spin down and place the tube back on the magnetic rack. Pipette off any residual ethanol. Allow the pellet to dry for ~30 seconds, but do not dry the pellet to the point of cracking.

Remove the tube from the magnetic rack and resuspend the pellet in 35 µl nuclease-free water by gently flicking.

Incubate for 10 minutes at 37°C. Every 2 minutes, agitate the sample by gently flicking for 10 seconds to encourage DNA elution.

Pellet the beads on a magnetic rack until the eluate is clear and colourless.

Remove and retain 35 µl of eluate into a clean 1.5 ml Eppendorf DNA LoBind tube.

CHECKPOINT

Quantify 1 µl of eluted sample using a Qubit fluorometer.

最終ステップ

Take forward the dual barcoded DNA library to the adapter ligation and clean-up step. However, at this point it is also possible to store the library at 4°C overnight.

8. Adapter ligation and clean-up

材料
  • Long Fragment Buffer (LFB)
  • Short Fragment Buffer (SFB)
  • Elution Buffer (EB)
  • Native Adapter (NA)
  • AMPure XP Beads (AXP)

消耗品
  • NEBNext® Quick Ligation Module (NEB, E6056)
  • 1.5 ml Eppendorf DNA LoBind tubes
  • Qubit™ Assay Tubes (Invitrogen, Q32856)
  • Qubit dsDNA HS Assay Kit (ThermoFisher, cat # Q32851)

装置
  • Microfuge
  • Magnetic rack
  • Vortex mixer
  • Hula mixer (gentle rotator mixer)
  • Thermal cycler
  • P1000 pipette and tips
  • P200 pipette and tips
  • P100 pipette and tips
  • P20 pipette and tips
  • P10 pipette and tips
  • Ice bucket with ice
  • Qubit fluorometer (or equivalent for QC check)
重要

The Native Adapter (NA) used in this kit and protocol is not interchangeable with other sequencing adapters.

Prepare the NEBNext Quick Ligation Reaction Module according to the manufacturer's instructions, and place on ice:

  1. Thaw the reagents at room temperature.

  2. Spin down the reagent tubes for 5 seconds.

  3. Ensure the reagents are fully mixed by performing 10 full volume pipette mixes. Note: Do NOT vortex the Quick T4 DNA Ligase.

The NEBNext Quick Ligation Reaction Buffer (5x) may have a little precipitate. Allow the mixture to come to room temperature and pipette the buffer up and down several times to break up the precipitate, followed by vortexing the tube for several seconds to ensure the reagent is thoroughly mixed.

重要

Do not vortex the Quick T4 DNA Ligase.

Spin down the Native Adapter (NA) and Quick T4 DNA Ligase, pipette mix and place on ice.

Thaw the Elution Buffer (EB) at room temperature and mix by vortexing. Then spin down and place on ice.

重要

Depending on the wash buffer (LFB or SFB) used, the clean-up step after adapter ligation is designed to either enrich for DNA fragments of >3 kb, or purify all fragments equally.

  • To enrich for DNA fragments of 3 kb or longer, use Long Fragment Buffer (LFB)
  • To retain DNA fragments of all sizes, use Short Fragment Buffer (SFB)

Thaw either Long Fragment Buffer (LFB) or Short Fragment Buffer (SFB) at room temperature and mix by vortexing. Then spin down and place on ice.

In a 1.5 ml Eppendorf LoBind tube, mix in the following order:

Between each addition, pipette mix 10 - 20 times.

Reagent Volume
Pooled barcoded libraries 30 µl
Native Adapter (NA) 5 µl
NEBNext Quick Ligation Reaction Buffer (5X) 10 µl
Quick T4 DNA Ligase 5 µl
Total 50 µl

Thoroughly mix the reaction by gently pipetting and briefly spinning down.

Incubate the reaction for 20 minutes at room temperature.

重要

The next clean-up step uses Long Fragment Buffer (LFB) or Short Fragment Buffer (SFB) rather than 80% ethanol to wash the beads. The use of ethanol will be detrimental to the sequencing reaction.

Resuspend the AMPure XP Beads (AXP) by vortexing.

Add 20 µl of resuspended AMPure XP Beads (AXP) to the reaction and mix by pipetting.

Incubate on a Hula mixer (rotator mixer) for 10 minutes at room temperature.

Spin down the sample and pellet on the magnetic rack. Keep the tube on the magnet and pipette off the supernatant.

Wash the beads by adding either 125 μl Long Fragment Buffer (LFB) or Short Fragment Buffer (SFB). Flick the beads to resuspend, spin down, then return the tube to the magnetic rack and allow the beads to pellet. Remove the supernatant using a pipette and discard.

Repeat the previous step.

Spin down and place the tube back on the magnet. Pipette off any residual supernatant.

Remove the tube from the magnetic rack and resuspend pellet in 15 µl Elution Buffer (EB).

Spin down and incubate for 10 minutes at 37°C. Every 2 minutes, agitate the sample by gently flicking for 10 seconds to encourage DNA elution.

Pellet the beads on a magnet until the eluate is clear and colourless, for at least 1 minute.

Remove and retain 15 µl of eluate containing the DNA library into a clean 1.5 ml Eppendorf DNA LoBind tube.

Dispose of the pelleted beads

CHECKPOINT

Quantify 1 µl of eluted sample using a Qubit fluorometer.

Depending on your DNA library fragment size, prepare your final library in 12 µl of Elution Buffer (EB).

Fragment library length Flow cell loading amount
Very short (<1 kb) 100 fmol
Short (1-10 kb) 35–50 fmol
Long (>10 kb) 300 ng

Note: If the library yields are below the input recommendations, load the entire library.

If required, we recommend using a mass to mol calculator such as the NEB calculator.

最終ステップ

The prepared library is used for loading onto the flow cell. Store the library on ice or at 4°C until ready to load.

ヒント

Library storage recommendations

We recommend storing libraries in Eppendorf DNA LoBind tubes at 4°C for short-term storage or repeated use, for example, re-loading flow cells between washes. For single use and long-term storage of more than 3 months, we recommend storing libraries at -80°C in Eppendorf DNA LoBind tubes.

オプショナルステップ

If quantities allow, the library may be diluted in Elution Buffer (EB) for splitting across multiple flow cells.

Depending on how many flow cells the library will be split across, more Elution Buffer (EB) than what is supplied in the kit will be required.

9. Priming and loading the SpotON flow cell

材料
  • Flow Cell Flush (FCF)
  • Flow Cell Tether (FCT)
  • Library Solution (LIS)
  • Library Beads (LIB)
  • Sequencing Buffer (SB)

消耗品
  • 1.5 ml Eppendorf DNA LoBind tubes
  • MinION and GridION Flow Cell
  • Nuclease-free water (e.g. ThermoFisher, AM9937)
  • Bovine Serum Albumin (BSA) (50 mg/ml) (e.g Invitrogen™ UltraPure™ BSA 50 mg/ml, AM2616)

装置
  • MinION or GridION device
  • MinION and GridION Flow Cell Light Shield
  • P1000 pipette and tips
  • P100 pipette and tips
  • P20 pipette and tips
  • P10 pipette and tips
重要

Please note, this kit is only compatible with R10.4.1 flow cells (FLO-MIN114).

ヒント

Priming and loading a flow cell

We recommend all new users watch the 'Priming and loading your flow cell' video before your first run.

Using the Library Solution

For most sequencing experiments, use the Library Beads (LIB) for loading your library onto the flow cell. However, for viscous libraries it may be difficult to load with the beads and may be appropriate to load using the Library Solution (LIS).

Thaw the Sequencing Buffer (SB), Library Beads (LIB) or Library Solution (LIS, if using), Flow Cell Tether (FCT) and Flow Cell Flush (FCF) at room temperature before mixing by vortexing. Then spin down and store on ice.

重要

For optimal sequencing performance and improved output on MinION R10.4.1 flow cells (FLO-MIN114), we recommend adding Bovine Serum Albumin (BSA) to the flow cell priming mix at a final concentration of 0.2 mg/ml.

Note: We do not recommend using any other albumin type (e.g. recombinant human serum albumin).

To prepare the flow cell priming mix with BSA, combine Flow Cell Flush (FCF) and Flow Cell Tether (FCT), as directed below. Mix by pipetting at room temperature.

Note: We are in the process of reformatting our kits with single-use tubes into a bottle format. Please follow the instructions for your kit format.

Single-use tubes format: Add 5 µl Bovine Serum Albumin (BSA) at 50 mg/ml and 30 µl Flow Cell Tether (FCT) directly to a tube of Flow Cell Flush (FCF).

Bottle format: In a suitable tube for the number of flow cells, combine the following reagents:

Reagent Volume per flow cell
Flow Cell Flush (FCF) 1,170 µl
Bovine Serum Albumin (BSA) at 50 mg/ml 5 µl
Flow Cell Tether (FCT) 30 µl
Total volume 1,205 µl

Open the MinION or GridION device lid and slide the flow cell under the clip. Press down firmly on the flow cell to ensure correct thermal and electrical contact.

Flow Cell Loading Diagrams Step 1a

Flow Cell Loading Diagrams Step 1b

オプショナルステップ

Complete a flow cell check to assess the number of pores available before loading the library.

This step can be omitted if the flow cell has been checked previously.

See the flow cell check instructions in the MinKNOW protocol for more information.

Slide the flow cell priming port cover clockwise to open the priming port.

Flow Cell Loading Diagrams Step 2

重要

Take care when drawing back buffer from the flow cell. Do not remove more than 20-30 µl, and make sure that the array of pores are covered by buffer at all times. Introducing air bubbles into the array can irreversibly damage pores.

After opening the priming port, check for a small air bubble under the cover. Draw back a small volume to remove any bubbles:

  1. Set a P1000 pipette to 200 µl
  2. Insert the tip into the priming port
  3. Turn the wheel until the dial shows 220-230 µl, to draw back 20-30 µl, or until you can see a small volume of buffer entering the pipette tip

Note: Visually check that there is continuous buffer from the priming port across the sensor array.

Flow Cell Loading Diagrams Step 03 V5

Load 800 µl of the priming mix into the flow cell via the priming port, avoiding the introduction of air bubbles. Wait for five minutes. During this time, prepare the library for loading by following the steps below.

Flow Cell Loading Diagrams Step 04 V5

Thoroughly mix the contents of the Library Beads (LIB) by pipetting.

重要

The Library Beads (LIB) tube contains a suspension of beads. These beads settle very quickly. It is vital that they are mixed immediately before use.

We recommend using the Library Beads (LIB) for most sequencing experiments. However, the Library Solution (LIS) is available for more viscous libraries.

In a new 1.5 ml Eppendorf DNA LoBind tube, prepare the library for loading as follows:

Reagent Volume per flow cell
Sequencing Buffer (SB) 37.5 µl
Library Beads (LIB) mixed immediately before use, or Library Solution (LIS), if using 25.5 µl
DNA library 12 µl
Total 75 µl

Complete the flow cell priming:

  1. Gently lift the SpotON sample port cover to make the SpotON sample port accessible.
  2. Load 200 µl of the priming mix into the flow cell priming port (not the SpotON sample port), avoiding the introduction of air bubbles.

Flow Cell Loading Diagrams Step 5

Flow Cell Loading Diagrams Step 06 V5

Mix the prepared library gently by pipetting up and down just prior to loading.

Add 75 μl of the prepared library to the flow cell via the SpotON sample port in a dropwise fashion. Ensure each drop flows into the port before adding the next.

Flow Cell Loading Diagrams Step 07 V5

Gently replace the SpotON sample port cover, making sure the bung enters the SpotON port and close the priming port.

Step 8 update

Flow Cell Loading Diagrams Step 9

重要

Install the light shield on your flow cell as soon as library has been loaded for optimal sequencing output.

We recommend leaving the light shield on the flow cell when library is loaded, including during any washing and reloading steps. The shield can be removed when the library has been removed from the flow cell.

Place the light shield onto the flow cell, as follows:

  1. Carefully place the leading edge of the light shield against the clip. Note: Do not force the light shield underneath the clip.

  2. Gently lower the light shield onto the flow cell. The light shield should sit around the SpotON cover, covering the entire top section of the flow cell.

J2264 - Light shield animation Flow Cell FAW optimised

注意

The MinION Flow Cell Light Shield is not secured to the flow cell and careful handling is required after installation.

最終ステップ

Close the device lid and set up a sequencing run on MinKNOW.

10. Data acquisition and basecalling

Overview of nanopore data analysis

For a full overview of nanopore data analysis, which includes options for basecalling and post-basecalling analysis, please refer to the Data Analysis document.

How to start sequencing

The sequencing device control, data acquisition and real-time basecalling are carried out by the MinKNOW software. Please ensure MinKNOW is installed on your computer or device. Instructions for this can be found in the MinKNOW protocol.

MinKNOW settings for real-time basecalling:

We recommend setting up real-time basecalling as follows:

Real-time basecalling

Follow the instructions in the MinKNOW protocol beginning from the "Starting a sequencing run" section for your device until the end of the "Completing a MinKNOW run" section.

Select Ligation Sequencing Kit (SQK-LSK114) in "Kit selection". The barcoding option will be unavailable as default. Other parameters can be kept at their default settings.

Picture1

Post-run barcode demultiplexing:

We currently recommend performing demultiplexing for the dual barcoding protocol post-sequencing using the Guppy software.

For a complete guide please refer to our Guppy protocol

Barcode demultiplexing in Guppy for dual barcoding:

To demultiplex your reads by barcode, use the dual barcoding configuration file, specifying the barcode kit as "EXP-DUAL00":

guppy_barcoder --input_path <folder containing FASTQ and/or FASTA files> --save_path <output folder> --config configuration_dual.cfg --barcode_kits "EXP-DUAL00"

For more information and instructions on how to use Guppy, please refer to the relevant sections of the protocol:

Barcode demultiplexing Quick Start

11. Downstream analysis

Post-basecalling analysis

There are several options for further analysing your basecalled data:

1. EPI2ME workflows

For in-depth data analysis, Oxford Nanopore Technologies offers a range of bioinformatics tutorials and workflows available in EPI2ME, which are available in the EPI2ME section of the Community. The platform provides a vehicle where workflows deposited in GitHub by our Research and Applications teams can be showcased with descriptive texts, functional bioinformatics code and example data.

2. Research analysis tools

Oxford Nanopore Technologies' Research division has created a number of analysis tools, that are available in the Oxford Nanopore GitHub repository. The tools are aimed at advanced users, and contain instructions for how to install and run the software. They are provided as-is, with minimal support.

3. Community-developed analysis tools

If a data analysis method for your research question is not provided in any of the resources above, please refer to the resource centre and search for bioinformatics tools for your application. Numerous members of the Nanopore Community have developed their own tools and pipelines for analysing nanopore sequencing data, most of which are available on GitHub. Please be aware that these tools are not supported by Oxford Nanopore Technologies, and are not guaranteed to be compatible with the latest chemistry/software configuration.

12. Flow cell reuse and returns

材料
  • Flow Cell Wash Kit (EXP-WSH004)

After your sequencing experiment is complete, if you would like to reuse the flow cell, please follow the Flow Cell Wash Kit protocol and store the washed flow cell at 2-8°C.

The Flow Cell Wash Kit protocol is available on the Nanopore Community.

ヒント

We recommend you to wash the flow cell as soon as possible after you stop the run. However, if this is not possible, leave the flow cell on the device and wash it the next day.

Alternatively, follow the returns procedure to flush out the flow cell ready to send back to Oxford Nanopore.

Instructions for returning flow cells can be found here.

Note: All flow cells must be flushed with deionised water before returning the product.

重要

If you encounter issues or have questions about your sequencing experiment, please refer to the Troubleshooting Guide that can be found in the online version of this protocol.

13. Issues during DNA/RNA extraction and library preparation for Kit 14

Below is a list of the most commonly encountered issues, with some suggested causes and solutions.

We also have an FAQ section available on the Nanopore Community Support section.

If you have tried our suggested solutions and the issue still persists, please contact Technical Support via email (support@nanoporetech.com) or via LiveChat in the Nanopore Community.

Low sample quality

Observation Possible cause Comments and actions
Low DNA purity (Nanodrop reading for DNA OD 260/280 is <1.8 and OD 260/230 is <2.0–2.2) The DNA extraction method does not provide the required purity The effects of contaminants are shown in the Contaminants document. Please try an alternative extraction method that does not result in contaminant carryover.

Consider performing an additional SPRI clean-up step.
Low RNA integrity (RNA integrity number <9.5 RIN, or the rRNA band is shown as a smear on the gel) The RNA degraded during extraction Try a different RNA extraction method. For more info on RIN, please see the RNA Integrity Number document. Further information can be found in the DNA/RNA Handling page.
RNA has a shorter than expected fragment length The RNA degraded during extraction Try a different RNA extraction method. For more info on RIN, please see the RNA Integrity Number document. Further information can be found in the DNA/RNA Handling page.

We recommend working in an RNase-free environment, and to keep your lab equipment RNase-free when working with RNA.

Low DNA recovery after AMPure bead clean-up

Observation Possible cause Comments and actions
Low recovery DNA loss due to a lower than intended AMPure beads-to-sample ratio 1. AMPure beads settle quickly, so ensure they are well resuspended before adding them to the sample.

2. When the AMPure beads-to-sample ratio is lower than 0.4:1, DNA fragments of any size will be lost during the clean-up.
Low recovery DNA fragments are shorter than expected The lower the AMPure beads-to-sample ratio, the more stringent the selection against short fragments. Please always determine the input DNA length on an agarose gel (or other gel electrophoresis methods) and then calculate the appropriate amount of AMPure beads to use. SPRI cleanup
Low recovery after end-prep The wash step used ethanol <70% DNA will be eluted from the beads when using ethanol <70%. Make sure to use the correct percentage.

14. Issues during the sequencing run for Kit 14

Below is a list of the most commonly encountered issues, with some suggested causes and solutions.

We also have an FAQ section available on the Nanopore Community Support section.

If you have tried our suggested solutions and the issue still persists, please contact Technical Support via email (support@nanoporetech.com) or via LiveChat in the Nanopore Community.

Fewer pores at the start of sequencing than after Flow Cell Check

Observation Possible cause Comments and actions
MinKNOW reported a lower number of pores at the start of sequencing than the number reported by the Flow Cell Check An air bubble was introduced into the nanopore array After the Flow Cell Check it is essential to remove any air bubbles near the priming port before priming the flow cell. If not removed, the air bubble can travel to the nanopore array and irreversibly damage the nanopores that have been exposed to air. The best practice to prevent this from happening is demonstrated in this video.
MinKNOW reported a lower number of pores at the start of sequencing than the number reported by the Flow Cell Check The flow cell is not correctly inserted into the device Stop the sequencing run, remove the flow cell from the sequencing device and insert it again, checking that the flow cell is firmly seated in the device and that it has reached the target temperature. If applicable, try a different position on the device (GridION/PromethION).
MinKNOW reported a lower number of pores at the start of sequencing than the number reported by the Flow Cell Check Contaminations in the library damaged or blocked the pores The pore count during the Flow Cell Check is performed using the QC DNA molecules present in the flow cell storage buffer. At the start of sequencing, the library itself is used to estimate the number of active pores. Because of this, variability of about 10% in the number of pores is expected. A significantly lower pore count reported at the start of sequencing can be due to contaminants in the library that have damaged the membranes or blocked the pores. Alternative DNA/RNA extraction or purification methods may be needed to improve the purity of the input material. The effects of contaminants are shown in the Contaminants Know-how piece. Please try an alternative extraction method that does not result in contaminant carryover.

MinKNOW script failed

Observation Possible cause Comments and actions
MinKNOW shows "Script failed"
Restart the computer and then restart MinKNOW. If the issue persists, please collect the MinKNOW log files and contact Technical Support. If you do not have another sequencing device available, we recommend storing the flow cell and the loaded library at 4°C and contact Technical Support for further storage guidance.

Pore occupancy below 40%

Observation Possible cause Comments and actions
Pore occupancy <40% Not enough library was loaded on the flow cell 10–20 fmol of good quality library can be loaded on to a MinION/GridION flow cell. Please quantify the library before loading and calculate mols using tools like the Promega Biomath Calculator, choosing "dsDNA: µg to pmol"
Pore occupancy close to 0 The Native Barcoding Kit was used, and ethanol was used instead of LFB or SFB at the wash step after sequencing adapter ligation Ethanol can denature the motor protein on the sequencing adapters. Make sure the LFB or SFB buffer was used after ligation of sequencing adapters.
Pore occupancy close to 0 No tether on the flow cell Tethers are adding during flow cell priming (FCT tube). Make sure FCT was added to FCF before priming.

Shorter than expected read length

Observation Possible cause Comments and actions
Shorter than expected read length Unwanted fragmentation of DNA sample Read length reflects input DNA fragment length. Input DNA can be fragmented during extraction and library prep.

1. Please review the Extraction Methods in the Nanopore Community for best practice for extraction.

2. Visualise the input DNA fragment length distribution on an agarose gel before proceeding to the library prep. DNA gel2 In the image above, Sample 1 is of high molecular weight, whereas Sample 2 has been fragmented.

3. During library prep, avoid pipetting and vortexing when mixing reagents. Flicking or inverting the tube is sufficient.

Large proportion of unavailable pores

Observation Possible cause Comments and actions
Large proportion of unavailable pores (shown as blue in the channels panel and pore activity plot)

image2022-3-25 10-43-25 The pore activity plot above shows an increasing proportion of "unavailable" pores over time.
Contaminants are present in the sample Some contaminants can be cleared from the pores by the unblocking function built into MinKNOW. If this is successful, the pore status will change to "sequencing pore". If the portion of unavailable pores stays large or increases:

1. A nuclease flush using the Flow Cell Wash Kit (EXP-WSH004) can be performed, or
2. Run several cycles of PCR to try and dilute any contaminants that may be causing problems.

Large proportion of inactive pores

Observation Possible cause Comments and actions
Large proportion of inactive/unavailable pores (shown as light blue in the channels panel and pore activity plot. Pores or membranes are irreversibly damaged) Air bubbles have been introduced into the flow cell Air bubbles introduced through flow cell priming and library loading can irreversibly damage the pores. Watch the Priming and loading your flow cell video for best practice
Large proportion of inactive/unavailable pores Certain compounds co-purified with DNA Known compounds, include polysaccharides, typically associate with plant genomic DNA.

1. Please refer to the Plant leaf DNA extraction method.
2. Clean-up using the QIAGEN PowerClean Pro kit.
3. Perform a whole genome amplification with the original gDNA sample using the QIAGEN REPLI-g kit.
Large proportion of inactive/unavailable pores Contaminants are present in the sample The effects of contaminants are shown in the Contaminants Know-how piece. Please try an alternative extraction method that does not result in contaminant carryover.

Reduction in sequencing speed and q-score later into the run

Observation Possible cause Comments and actions
Reduction in sequencing speed and q-score later into the run Fast fuel consumption is typically seen in Kit 9 chemistry (e.g. SQK-LSK109) when the flow cell is overloaded with library. Please see the appropriate protocol for your DNA library to find the recommendation. Add more fuel to the flow cell by following the instructions in the MinKNOW protocol. In future experiments, load lower amounts of library to the flow cell.

Temperature fluctuation

Observation Possible cause Comments and actions
Temperature fluctuation The flow cell has lost contact with the device Check that there is a heat pad covering the metal plate on the back of the flow cell. Re-insert the flow cell and press it down to make sure the connector pins are firmly in contact with the device. If the problem persists, please contact Technical Services.

Failed to reach target temperature

Observation Possible cause Comments and actions
MinKNOW shows "Failed to reach target temperature" The instrument was placed in a location that is colder than normal room temperature, or a location with poor ventilation (which leads to the flow cells overheating) MinKNOW has a default timeframe for the flow cell to reach the target temperature. Once the timeframe is exceeded, an error message will appear and the sequencing experiment will continue. However, sequencing at an incorrect temperature may lead to a decrease in throughput and lower q-scores. Please adjust the location of the sequencing device to ensure that it is placed at room temperature with good ventilation, then re-start the process in MinKNOW. Please refer to this FAQ for more information on MinION temperature control.

Last updated: 11/15/2024

Document options