MinION Mk1C user manual

Overview

This user manual will enable the user of the MinION Mk1C to:

  • Power up and down the device
  • Learn the shortcuts enabled by touchscreen in the MinION Mk1C Graphical User Interface (GUI)
  • Use the Host Settings for system software configuration
  • Access the device remotely for control of MinKNOW and to transfer data on Windows, macOS and Linux computers
  • Access the device through SecureShell/the Terminal for data transfer and device configuration

Our device FAQs are located here.

For Research Use Only

Document version: MKC_2005_v1_revAG_27Nov2019

1. Device information

The MinION Mk1C

The MinION Mk1C combines the real-time, rapid, portable sequencing of a MinION Mk1B with a Graphical Processing Unit (GPU) and a high-resolution screen. The device applies the technology of the MinION Mk1B with the basecalling power of the MinIT and the additional features of smartphone, such as a cellular modem and touch screen.

component part image for mk 1c - white border small

The hardware

Component Specification
Model Number MIN-101C
Size and weight H 32 mm x L 142 mm x W 118 mm; 440 g
Power • Supplied with a 6.3-19.6 VDC power supply
• Max rated current 10 A
• Max rated power 60 W
Compute spec 1 TB SSD Storage, 8 GB RAM, GPU embedded analysis accelerator (ARM processor 6 cores, 256 core GPU)
Connections • 1x USB 2.0
• 1x eSATA (currently non-functional)
• 1x microSD card
• 1x 1 GB Ethernet
• Wi-Fi connectivity
Pre-loaded software Linux OS, MinKNOW and Dorado
Security Kensington Nano Security Slot
Environmental conditions • Designed to sequence at +10°C to +30°C*
• Do not cover vents on the sides of the device
• Indoor use
• Altitude up to 2,000 metres
• Maximum relative humidity 80% for temperatures up to 31°C, decreasing linearly to 50% relative humidity at 40°C
• Applicable pollution degree 2 in the intended environment

* Functional range of electronics +5°C to +40°C

FAQs

Our MinION Mk1C FAQs are located here.

2. What's in the box

What's in the box

The MinION Mk1C is shipped with:

  • Configuration Test Cell (CTC)
  • Power adapter
  • Plugs compatible with UK, US, EU and Australian sockets
  • Quick Start Guide

MinION Mk1C transparent packaging power adapter UK EU US AU plugs (2)

MinION MK1C transparent square cable power adapter plugs UK EU US AU (2)

The components of the MinION Mk1C are shown below.

Mk1C annotated

Touch screen The touch screen allows the user to interact with the MinION Mk1C software directly on the device. LED lights There are five LED lights to the right of the touch screen. All five LEDs illuminate white when the device starts up and then cycle as the system software loads. The cycling indicates that the device is working. When a USB is inserted into the MinION Mk1C, all LEDs illuminate and flash three times before the normal cyclical pattern resumes. Lid The lid can be flipped open to access the flow cell housing. Clip The clip holds the flow cell or Configuration Test Cell securely in place. Configuration Test Cell (CTC) The CTC is used during the hardware check to ensure that the communication between the device and the flow cell is working correctly.

The connections and ports available on the MinION Mk1C are shown below.

Mk1C labelled

3. Powering the device

The AC/DC adapter power specification

The power requirements for the MinION Mk1C using the AC/DC adapter supplied with the device are detailed below.

Power supply Power input Power output
AC/DC adapter Standard Mains input:
100-240 VAC
50/60 Hz
1.5 A
Output: 6.3-19.6 VDC
Max rated current 10 A
Max rated power 60 W

Estimated power dissipation:
20-25 W at room temperature

The MinION Mk1C is compatible with portable DC power supplies.

Oxford Nanopore Technologies does not supply a portable power supply for this purpose and as such it is not officially supported. If you wish to make use of a portable power supply, then it must be compatible with the below specification.

DC power pack requirements

Power supply Power input Power output
DC power pack Standard Mains input for charging:
100-240 VAC
50/60 Hz
Estimated power dissipation:
20-25 W at room temperature

Fit the appropriate pin-plug into the base unit of the AC/DC adapter.

power adapter plufg ends

Insert the barrel pin into the MinION Mk1C power port.

Plug for Mk 1C - how to assemble-plugging barrel adapter into socket

Switch on the AC Mains.

Press the power button on the device to switch it on.

Power button red box

Once the MinION Mk1C is powered on, the power button will illuminate blue and the device fans will sound.

On switching on the device for the first time, users will be taken through the onboarding flow for setting up the device. Once the MinION Mk1C is set up, users will need to log into their MinKNOW accounts when switching on their device if the onboarding flow has already been completed.

If you experience login issues, contact Technical Support via email (support@nanoporetech.com) or via LiveChat in the Nanopore Community.

To log in, you must be connected to the internet.

login page

For more information see the Onboarding process section in this document.

4. Onboarding process

Onboarding process for new devices

The onboarding process is available when a MinION Mk1C is first switched on to guide users through setting up a network connection, logging into MinKNOW, enabling remote access and running a hardware check.

Update to the latest version of MinKNOW for the MinION Mk1C, which supports the latest chemistry, bug fixes and performance enhancements.

Please visit our Oxford Nanopore FAQ and Help Centre if you are having difficulties updating MinKNOW on the MinION Mk1C.

Select the language for the device.

Onboarding lang

Choose the network connection to connect to the internet.

We strongly recommend users to connect their device to the internet to allow them to log into their MinKNOW account.

Onboarding network

There are two options:

  1. Connect using WiFi: click WiFi and select the network to join and enter password, if prompted.

  2. Connect using Ethernet: plug-in the Ethernet cable and click Ethernet.

Note: If your network configuration requires you to use a static IP address, click Offline and configure the network through the main Network Settings page.

Log into your MinKNOW account using your Nanopore account credentials.

Onboarding login

Configure remote access from the options displayed.

These settings can be altered again using the Host settings after the onboarding flow has ended.

Onboarding remote access

Complete a hardware check to ensure the device is working correctly for sequencing experiments.

A hardware check only needs to be completed once. However, it can be skipped for users who will not be using their device for sequencing.

Onboarding hardware

Navigate to the Device Settings to verify that the date and time are correct. Having the correct time is important to successfully establishing a connection to a network.

5. General connection information

Connection

There are several options to connect to your MinION Mk1C. These are:

General Options

  • If you would like to use the sequencing software (MinKNOW) directly on the MinION Mk1C device: Log into MinKNOW using the device GUI and select the MinION Mk1C device on the Connection Manager.
  • If you would like to connect to the MinION Mk1C remotely using a computer: Download MinKNOW software and select the MinKNOW icon shortcut and log into MinKNOW. Select the MinION Mk1C device on the Connection Manager to remotely connect.

Advanced users

MinKNOW and data access with your MinION Mk1C

MinION Mk1C sequences and basecalls data via the integrated sequencing device and Oxford Nanopore Technologies basecaller - Dorado - in MinKNOW. POD5, FASTQ and BAM files are output and stored locally on the MinION Mk1C in the Data folder.

This folder and the output files can be accessed through File Manager in MinKNOW. For more information, please see the Data management section of this document.

6. Changing the default password and hostname

Navigate to the host settings to open 'Device settings'.

Click 'Reset password' in the System box.

Mk1C password

In the 'Change password' dialogue box, enter your current password, followed by the new password.

Password change dialogue

Click Reset Password to confirm password change.

(Advanced) Changing the default hostname

  • If you need to update the hostname of the device, run the following command and replace the <new_hostname> portion with your desired hostname (omitting the angle bracket characters <>):

sudo -S hostnamectl set-hostname <new_hostname>

  • Power cycle the device by shutting down with the following command, then press the physical power button to turn the device back on:

sudo shutdown -h now

7. Network settings and connection

Network settings display internet connectivity information. The MinION Mk1C can be connected via Ethernet or Wi-Fi.

Both options are supported, however if Ethernet is available, we recommend using Ethernet because the connection will likely be quicker and more reliable.

Ethernet

To connect using Ethernet:

  1. Connect a Ethernet cable to the device
  2. Navigate to network settings in host settings to confirm connection. IP configuration information will appear when connected.

Mk1c 1

Wi-Fi

To connect using Wi-Fi, navigate to network settings and toggle Wi-Fi on. Select a network from the Available Networks panel to connect to the internet.

Network settings 2

Select Advanced to open Advanced Wi-Fi Configurations. From here, all available networks can be viewed, connected or removed.

Network settings 4

IMPORTANT

Please note that Wi-Fi connections with captive portals (a web page that the user is required to view and enter login details before access is granted) cannot be used with the MinION Mk1C.

Configuring static IP address

The static IP address can be configured through MinKNOW UI for both WiFI and Ethernet. Below is an example for Ethernet.

  1. Click Edit in "IP configuration" to open "Edit the IP address configuration" dialogue box. Mk1c 1
  2. Select either DHCP or Static tab and click Confirm to save. Mk1c 3
  3. Once the IP address configuration has changed successfully, click Finish to close the dialogue box. Mk1c 4

Hotspot can be activated to allow users to connect to a larger screen.

This can be used to connect to the MinION Mk1C in places where there is no network infrastructure, e.g. in the field. To enable hotspot, select the toggle to ON. When active, the device hotspot name will be displayed.

image (62)

Hotspot configs can be edited to select a channel if required for congested networks where there may be competition for channels.

  1. Navigate to the network settings
  2. For hotspot, select the toggle to ON
  3. Click 'Configure hotspot'
  4. Select channel using the dropdown menu
  5. Enter new password
  6. Click 'Confirm'

The hotspot should be confirmed as specified in the hotspot configs.

wifi hotspot mk1c

To connect the MinION Mk1C Wi-Fi router to a Windows device:

  1. Locate the MinION Mk1C ID on the underside label of the device:

MinION Mk1C serial label

  1. Click on the MinION Mk1C ID under the Wi-Fi Connections Panel in the Control Panel or the options in the right side of the Task Bar

Login with the Security Key: WarmButterflyWings98

mc-110016 Wifi ID

  1. Access the MinION Mk1C by the File Explorer: Go to \\mc-XXXXXX (XXXXXX is the MinION Mk1C ID) in the Address Bar. You will be prompted for the network credentials on connecting to the MinION Mk1C through File Explorer.

To connect the MinION Mk1C router to a macOS device:

  1. Locate the MinION Mk1C ID on the underside label of the device:

MinION Mk1C serial label

  1. Click on the Wi-Fi icon on the right of Menu bar at the top of the screen:

Wi-Fi options white border 3.0 smaller

Identify the MC-XXXXXX ID in the Network Name under Network Panel (the final option under the Wi-Fi icon), which is show in the window below (left)‡. Selecting the MinION Mk1C ID and clicking Apply will present the Wi-Fi login pop-up. Use the password credentials shown below.

Login with the Password: WarmButterflyWings98

wifi log in for macOS

  1. Access the MinION Mk1C by the Network browser: Search for Network using Finder. Click on Server in the tab to lower left of the Search bar.

MC-XXXXXX (XXXXXX is the MinION Mk1C ID) will be displayed as a server option. Click the MC-XXXXXX icon, click Connect as under the Search bar and enter the network credentials.

connect as MAC

‡ When connecting to your device by SSH, it can be addressed in three different ways:

  1. Using DNS
  2. Using mDNS
  3. Using its IP address directly

In most organisations, central DNS servers are used. If your device is using a 'static' IP configuration, your IT department will have to create a DNS entry for it manually. If your device is using a dynamic (DHCP) IP configuration, the DNS entry will often be created automatically. To SSH into your device using DNS, set the hostname in your client to the device's hostname, e.g. MC-112345.

If DNS is not provided by your organisation or your broadband router, try using mDNS. When your device is connected to a network, it announces its presence to all other nearby devices such as your laptop or PC. If your laptop or PC is configured to recognise these announcements, you can SSH into your device using mDNS by setting the hostname in your client to the device's hostname followed by ".local", e.g. MC-112345.local.

If the above methods are not successful, use the devices IP address directly. The MinKNOW UI shows the IP addresses for each interface in the Settings - Network page. If your device has LAN and Wi-Fi interfaces, each will have a different IP address. To SSH into your device using its IP address, set the hostname in your client to the IP address, e.g. 10.20.30.41.

To connect the MinION Mk1C router to a Linux device:

  1. Locate the MinION Mk1C ID on the underside label of the device:

MinION Mk1C serial label

  1. Click on the Wi-Fi icon on the right of Menu bar at the top of the screen.

Wi-Fi options for Linux 3.0 white border

Identify and click on the MC-XXXXXX ID under the Wi-Fi network options icon (shown above). Enter the Wi-Fi password (shown below) in the pop-up Wi-Fi Network Authentication window:

Login with the Password: WarmButterflyWings98

Linux Wi-fi password - 2.0 white border

The following window will be presented when you have connected to the MinION Mk1C Wi-Fi router.

Wi-Fi connected for Linux 2.0 white border

  1. Access the MinION Mk1C using the Linux File Browser and enter the server address.

Open the File Browser (this can be found using the Search tool). Select Connect to server on the bottom of the list in the left panel. You will need to enter:

smb://mc-110013

8. Mount network drive

Mount drives

The MinION Mk1C contains 1 TB SSD storage, which provides an average capacity of roughly 50 Gbases, stored in .POD5, .fast5 and/or FASTQ format. Once the SSD fills up, you will need to move your sequencing data off the device to a networked storage location. Network drives for data storage can be mounted from either the Device Settings or the File Manager in the Network tab:

  • Device Settings: 68

  • File manager: 16

Click "Add network drive".

A modal will open. Click either SMB (Samba) or NFS (Network File System) shares.

Use SMB for connecting to a shared drive on a Windows server, and NFS for Linux. If your shared drive is on a macOS machine or NAS (Network-Attached Storage), you can use either SMB or NFS.

SMB: HS 2

NFS: HS 3

To mount SMB shares:

  1. Enter the host name or IP address of the SMB share. This is acquired from your network admin.
  2. Enter the drive on the network host that the user wants to share. Take note to start with a '/'. E.g. /data.
  3. The 'Mount Point' field will auto-populate with the host name and path to the network share. However, you can customise this. In the file manager, it will appear as: /data/network/mt-111111-data for example.
  4. Add a username.
  5. (Optional) Enter a password.
  6. (Optional) Enter a domain name.
  7. Click Mount.

Mount network drive 3

Providing the settings were correct, the mounted drive will be added to the 'Disk Management' overview. The user will be able to:

  • View the space information for the drive
  • Unmount the network drive
  • A link to jump to the mount point inside the file manager

To mount NFS shares:

  1. Enter the host name or IP address of the NFS share. Obtain this from your network admin.
  2. Enter the drive on the network host that the user wants to share. Take note to start with a '/'. E.g. /data.
  3. The 'Mount Point' field will auto-populate with the host name and path to the network share. However, you can customise this. In the file manager, it will appear as: /data/network/mt-111111-data for example.
  4. Click Mount.

Mount network drive 4

Providing the settings were correct, the mounted drive will be added to the 'Disk Management' overview. You will be able to:

  • View the space information for the drive
  • Unmount the network drive
  • A link to jump to the mount point inside the file manager

Click "Add network drive".

9. (Advanced) Connecting by the command-line

IMPORTANT

If you are uncertain about changing any network parameters on your device, you will need to contact your IT department.

Network configuration - connection to your MinION Mk1C

The MinION Mk1C has two interfaces that we support for data transfer. These are:

  • Ethernet port - direct to interface or LAN
  • Wi-Fi - direct to interface

Mk1C data transfer

Networking with MinION Mk1C. Blue lines indicate a connection via the internet. The black line indicates a wired connection. The computer on the LAN (B.) can be used to configure the device using parameters that are not available through the Admin interface. You can also configure the MinION Mk1C using the Wi-Fi connected device (A.), however network compatibility issues may inhibit the update.

Internet connection:

This configuration requires an ethernet connection to the Local Area Network (LAN) from the MinION Mk1C. The LAN is connected to the internet, which enables updating of MinKNOW on the device.

Interface connection:

An interface, such as a laptop/mobile device/computer, can be wirelessly connected to the MinION Mk1C. Alternatively, the MinION Mk1C may be accessed and controlled by a computer on the LAN.

Other configurations are also possible, however they are not supported by Oxford Nanopore Technologies:

  • Ethernet-desktop/laptop computer to control the MinION Mk1C - no internet
  • MinION Mk1C Wi-Fi (Client) to a router and desktop/laptop Wi-Fi to a router, which is itself connected to the internet
  • It is possible to adapt other data connections through the USB 2.0 port, however, Oxford Nanopore Technologies cannot advise or support this use.

Connecting by SSH/the terminal provides options to configure and access your MinION Mk1C computer and network.

Connecting by the command line via SSH or the terminal enables several options:

  • Access and transfer data on to a mounted USB-hard drive - for instructions, see the Data transfer via SFTP section of this user manual.
  • Change the output directory from /data
  • Change between static and DHCP (dynamic) IP address of the ports on the MinION Mk1C
  • Access to the logs

Windows

Option 1

If SSH is enabled in your Windows build, you can use the built-in Windows Command Prompt (CMD). Open a CMD prompt, then enter ssh minit@hostname (e.g. ssh minit@mc-111414):

ssh windows cmd 1

If it is your first time connecting, you will be asked to confirm the authenticity of the host. Enter yes to continue:

ssh windows cmd 2

Enter the password to establish an SSH connection. The default password is minit. Please note that characters will not appear on-screen as you type in the password.

ssh windows cmd 3

If the SSH connection is successfully established, the shell prompt will now read minit@hostname:~$

ssh windows cmd 4

Option 2

A third-party SSH client such as Bitvise (example images shown below) or Putty may be used. The following default options can be entered:

Host – device ID or IP address (e.g. MC-111414) Port – 22 Username – minit Initial method – password Password – minit

ssh windows bitvise 1

If a connection is successfully established, Bitvise will automatically open a terminal window and an SFTP browser. Alternatively, you can click on New terminal console or New SFTP window on the left-hand side:

ssh windows bitvise 2

Linux/macOS

On both Linux and macOS operating systems, you can use the built-in Terminal program.

Open a Terminal, then type ssh minit@hostname (e.g. ssh minit@mc-111414):

ssh linux macos 1

If it is your first time connecting, you will be asked to confirm the authenticity of the host. Enter yes to continue:

ssh linux macos 2

Enter the password to establish an SSH connection. The default password is minit. Please note that characters will not appear on-screen as you type in the password.

ssh linux macos 3

If the SSH connection is successfully established, the shell prompt will now read minit@hostname:~$

ssh linux macos 4

10. (Advanced) Network configuration - static vs dynamic IP

IMPORTANT

If you are uncertain about changing any network parameters on your device, you will need to contact your IT department.

Network configuration - Static vs dynamic IP

The LAN interface on the MinION Mk1C can operate in either DHCP (dynamic) mode or static mode. When in DHCP mode, the settings are acquired from the network’s DHCP server. Some networks do not have a DHCP server or you may wish to override the settings it provides. In such circumstances, the static mode can be used.

The static mode comes pre-configured with the following settings:

IP Address/Subnet Mask: 192.168.0.1/24
Default Gateway: 192.168.0.1
DNS Servers: 9.9.9.9, 8.8.8.8

To configure static IP addressing on the Ethernet port:

Here is an example of possible static settings that may be provided by your IT department:

IP Address: 10.20.30.40
Subnet Mask: 255.255.255.128
Default Gateway: 10.20.30.1
DNS Servers: 10.0.0.53, 10.0.1.53

To change the static profile to match these settings:
  1. Log in to your MinION Mk1C using SSH via the WiFi connection.
  2. If you are given the subnet mask in the form xxx.xxx.xxx.xxx, convert this to CIDR format, which is a number between 1 and 32 (common values are 16, 24 and 25). You can use the following website to convert from xxx.xxx.xxx.xxx to a CIDR value: https://docs.netgate.com/pfsense/en/latest/network/cidr.html
  3. Look up the xxx.xxx.xxx.xxx value in the first column and find the corresponding CIDR value in the second column. In the example above, 255.255.255.128 corresponds to a CIDR value of /25.
  4. Once you have this value, open a terminal window and enter the following commands:

sudo nmcli con down dhcp

sudo nmcli con down static

sudo nmcli con mod static ipv4.method manual ipv4.address 10.20.30.40/25 ipv4.gateway 10.20.30.1 ipv4.dns 10.0.0.53,10.0.1.53 ipv4.routes "" connection.autoconnect yes

sudo nmcli con up static

In some instances, these commands may give an error message, however this is not a problem and the error should be ignored. The addresses specified in the third command line are the values that need to be changed to match particular settings. Note that there is no space between the DNS server addresses.

Use the following command to verify that your settings are correctly applied:

sudo nmcli dev show eth0

To configure dynamic IP addressing on the Ethernet port:

Open a terminal window and enter the following commands:

  sudo nmcli con down static
  sudo nmcli con up dhcp

11. Hardware check

IMPORTANT

Hardware check

A hardware check must be performed on all new devices or when software has been updated. This uses the Configuration Test Cell(s) (CTC), that come pre-inserted into your device in place of flow cells.

Note: If using a Flongle Flow Cell, we recommend regularly checking the Flongle adapter by inserting the empty adapter and completing a hardware check. For checking the device or flow cell position, use a MinION CTC, even if a Flongle Flow Cell will be used for sequencing.

Navigate to the Start homepage and select the 'Hardware Check' option.

The hardware check page will open. Mk1C hardware 1

Select 'Start' on the Hardware Check page to start check.

Note: To check the Flongle adapter, insert the EMPTY adapter. To check device or position, insert only a MinION CTC.

50

You will be automatically navigated to the Sequencing Overview page.

A loading bar will be displayed under the flow cell during the check.

END OF STEP

The hardware check will complete after approximately one minute.

Hardware check pass is indicated by a green check icon.

An orange check icon is a fail.

  1. Flongle hardware check pass
  2. MinION hardware check pass
  3. Flongle hardware check fail

MinION hardware check

If the hardware check fails, remove and reinsert the CTC, and run a hardware check again. If the check fails for a second time, please contact Technical Support via email (support@nanoporetech.com).

12. Flow cell check

Flow cell check

We strongly recommend performing a flow cell check before loading a DNA or RNA library to assess the number of pores available.

Oxford Nanopore Technologies will replace any flow cell that falls below the warranty number of active pores within 12 weeks of purchase, provided that you report the results within two days of performing the flow cell check and you have followed the storage recommendations.

Flow cell Minimum number of active pores covered by warranty
Flongle Flow Cell (FLO-FLG001) 50
MinION/GridION Flow Cell 800
PromethION Flow Cell 5000

Navigate to the Start page and select 'Flow Cell Check' to open the flow cell check page.

Mk1C flow cell check

When you see the flow cell type and flow cell IDs recognised, click 'Start' to begin.

2

You will be automatically navigated to the Sequencing Overview page.

A loading bar will be displayed under the flow cell during the check.

Flow cell health indicators

The quality of the flow cell will be shown as one of the three outcomes:

  1. Yellow exclamation mark (Flongle flow cell): The number of sequencing pores is below warranty. Take the flow cell out of the device, re-insert it and run a flow cell check again. If the flow cell is still below warranty, contact support@nanoporetech.com

  2. Green tick (MinION flow cell): The number of sequencing pores is above warranty.

  3. Question mark (PromethION flow cell): A flow cell check has not been run on the flow cell during this MinKNOW session.

Flow cell check icon all devices

Note: The indicator of quality will only remain visible during a MinKNOW session. Once the MinKNOW session has ended, the status of the flow cell will be erased.

13. Sequencing and monitoring the experiment

For instructions on how to set up a sequencing experiment and how to monitor the progress of the experiment, refer to the MinKNOW protocol in the Nanopore Community.

To run a sequencing experiment, you will need your DNA or RNA sample purified and a sequencing library prepared. For more information about sample preparation, refer to the Prepare section of our Documentation.

You can access the MinKNOW protocol by following this link.

14. Data management

File manager

Data can be managed and transferred from the file manager tab on the host settings.

Navigate through the tabs to view the data stored:

  • Internal tab: Data stored on the connected sequencing device (MinION Mk1C, GridION, PromethION)
  • Removable tab: Data stored on a connected removable storage device e.g. USB drive
  • Network tab: Data stored on a connected network drive. The network drive must be mounted prior.

HS 6

To remotely access the device as a shared/network drive:

Navigate to device settings and switch the Share toggle on.

Picture16

A new window will open to create a password for security.

Picture17

Click Set password, and after a few seconds the share toggle will be active.

Data will now be able to be shared and accessible on different networks.

image (26)

Note: The image above is an example of sharing on a GridION. On MinION devices, 'minit' will appear and PromethION devices will show 'prom' when active.

TIP

To stop sharing, click the toggle to turn off.

Note: The password will be removed and must be reset when sharing is turned back on.

To manage data between the internal, removable and network tabs:

Switch the Select toggle on to open the greyed out options Copy, Move, Delete or Rename in the bottom right corner.

Picture18

Select a folder or file to use the options in the bottom right corner.

Picture19

TIP

To use Move and Delete, select a file and choose an option to either move or delete.

Click either Copy or Move to open a new dialogue box with a list of all the drives accessible to the device.

Note: a USB drive or SD card plugged in will appear under the Devices folder.

Picture20

Click on a drive to open and navigate through the file directory to choose destination. Click either Copy or Move to confirm action.

New directories can be created from the device GUI using the New Directory icon.

Click the icon in the bottom right corner to open Create New Directory and type in a directory name. Click Create.

Picture24

Disk management

Navigate to Device settings to view availabe disk space on the device.

Any peripherals plugged in, including USB, can be ejected using the Eject option.

A network drive can also be connected to the device by selecting Mount Network Drive and filing in the required credentials for connecting using a Samba (SMB) server or a Network File System (NFS).

Compatible disk formats: SD and USB on MinION Mk1C supports DOS FAT, Windows 95 FAT, exFAT, and Linux ext(2|3|4).

Please refer to the 'Mount Network Drive' section of this protocol for more information.

68

15. Data transfer via SFTP

To transfer data using this method, you will first need to connect to your MinION Mk1C using SSH or via the Terminal.

For instructions, see the (Advanced) connecting by the command-line section of this user manual.

Filezilla - A multi-operating system (OS) compatible software for data transfer

You can use software like Filezilla, which enables data transfer for mounted hard drives on Linux, Windows, and macOS. It may also transfer using the SFTP protocol which is available whenever an SSH connection can be established.

Filezilla can be downloaded for various OS to enable data transfer from your MinION Mk1C.

The following default options can be entered: Host – device ID or IP address (e.g. MC-111414) Username – minit Password – minit Port – 22

sftp filezilla 1

If it is your first time connecting, you will be asked to confirm the authenticity of the host. Check the box to Always trust this host, add this key to the cache and click OK to continue:

sftp filezilla 2

If the SFTP connection is successfully established, you will see the Mk1C file directory shown on the right-hand side of the UI. Please note that Filezilla defaults to the home directory (/home/minit):

sftp filezilla 3

You can navigate to the /data directory by entering /data into the Remote site field and clicking Enter:

sftp filezilla 4

16. Rebooting and shutting down

Either press the power button or navigate to the 'Device settings' in the Host settings and click either 'Shutdown' or 'Reboot'.

29

Wait for the dialogue box to open and click either 'Shut down' or 'Reboot'.

shutdown modal

TIP

The reboot will disconnect the MinION Mk1C from any device that is remotely connecting via Wi-Fi or SSH.

You will need to reconnect the MinION Mk1C to any device that was connected before the reboot using the procedure described in the Network settings and connection section of this document.

END OF STEP

The powering down sequence:

You will initially see some scrolling code. After ~10 seconds, the screen will go blank and the lights at the side of the screen will power down. It is safe to remove the power cable from the MinION Mk1C at this point, if shutting down.

17. Updating the software via the UI

Overview

MinKNOW is the software used on nanopore sequencing devices. The software is described in more detail in the MinKNOW Technical Document. The MinKNOW Protocol provides instructions on how to use the software.

IMPORTANT

Updating the software

Oxford Nanopore Technologies will post a release note when new software or updates go out via the Nanopore Community.

We urge users to update as soon as reasonably possible after the release has been made available.

Note: Please check the information for firewall access to the necessary IP addresses and ranges in this section.

Users will not be able to update their device if connected remotely or if a run is in progress.

Firewall settings for Oxford Nanopore Technologies' devices

The Oxford Nanopore Technologies software will require access to the AWS IP ranges currently listed here:

http://docs.aws.amazon.com/general/latest/gr/aws-ip-ranges.html

Access to the following IP addresses is also needed:

              178.79.175.200  
               96.126.99.215  

Plug an Ethernet cable into the Ethernet port and connect to the network.

A dialogue box will open when a new update is available.

Select Get Update to update the device software automatically.

Updates may be skipped. However, we recommend to update the device as soon as updates are available. Some updates will be mandatory to use the device and are unable to be skipped.

36

Note: For MinION Mk1B, clicking Get Update will open the Software Downloads page on the connected computer for you to download the updated MinION software.

You can also update the device from the Software page of the Host settings.

Navigate to Software in the Host settings and click Get Update to open the installer window.

An update button will only appear when a new version of the software is available to download.

Note: For MinION Mk1B devices, clicking Get Update will open the Software Downloads page on the connected computer for you to select and download the updated MinION software.

Software update 1

Select 'Get update'

Click Install Update in the installer window to download the new update.

A change log for the update is available in this dialogue box.

Installer 1

Wait for the software to update; this may take several minutes.

Installer 2

Check the software version number under 'Current version'.

18. (Advanced) Updating the software via SSH

Although MinION Mk1C software can be updated via the user interface, there are occasions where software updates have to be executed manually. This can also be done if the user prefers to update via the command line, since the UI does not give feedback during updates.

Always refer to the release notes for any version-specific commands that may be supplied.

Connect to the MinION Mk1C using SSH.

Open a terminal window, and update the list of available software using the following commands:

sudo apt clean
sudo apt update
sudo apt install ont-mk1c-release

These commands should complete successfully, without any errors or warnings. If you do see an error or warning message, check your MinION Mk1C is able to connect to the Internet. Run the following command to diagnose network connection problems:
sudo /opt/ont/mooneye/bin/ont-mooneye-check-network --verbose --diagnose

Once the commands have completed successfully without errors, please reboot the device using the command:

sudo reboot

To update all software (operating system and the MinKNOW software), enter the following command:

export DEBIAN_FRONTEND=noninteractive
sudo apt upgrade
OPTIONAL ACTION

Alternatively, to update just the MinKNOW software, enter the following command:

export DEBIAN_FRONTEND=noninteractive
sudo apt install ont-mk1c-release

If you are prompted to continue, type ‘y’ and press Enter.

Restart your MinION Mk1C following the update for all processes to complete successfully (this will not be prompted for).

19. OS update/device recovery using microSD cards

From March 2022, we are delivering microSD cards to customers to update their OS so that they can use future versions of MinKNOW (22.03 onwards) with the latest features.

If you meet the requirements below, you can update your device before the microSD card arrives.

Note: there is no guarantee this process will work, as not all vendors of microSD cards have been tested internally at Oxford Nanopore.

IMPORTANT

Requirements

  • Blank 16 GB+ microSD SDXC card
  • Computer with a microSD port/USB SD card adapter
  • Balena Etcher for Windows or Mac (https://www.balena.io/etcher/)

Find out the current version of Ubuntu on your MinION Mk1C.

If your MinION Mk1C serial number is greater than MC-112454, your device shipped with Ubuntu 18.04 (Bionic).

Alternatively, on the user interface navigate to Settings and then Software. Click on the ‘I’ icon next to the MinKNOW version: you should see either ubuntu-bionic or ubuntu-xenial.

Download the correct SD card image.

If your Mk1C is currently on Ubuntu Xenial, download the following image (with MinKNOW 21.11 included): https://nanoporetech.box.com/s/bsuzar12lvjp15qawz421t9zghk59jw0

Please visit our Help website for information on how to update MinKNOW on the MinION Mk1C after upgrading to Bionic (https://help.nanoporetech.com).

The recovery SD card for devices on Bionic is available here (latest image has MinKNOW 22.12.5 preinstalled): https://nanoporetech.box.com/s/yftsu9fr9ukqilairdjx8xadtr1u5ygl

If your MinION Mk1C is already on Bionic, you will need to update to the more stable version if you are unable to update to MinKNOW 22.03.

Complete the factory restore.

Power down the Mk1C. Insert the microSD card and power on your Mk1C. After 30 seconds you should see a System Recovery pop-up screen:

recover screenshot

Select Factory Reset to re-image the device with the software on the microSD card.

Follow the on-screen instructions to complete the OS update.

If no screen is observed, it is likely you have downloaded the wrong microSD card OS image, so check your OS version again. Alternatively, your device needs to perform some pre-requisite actions, see the Troubleshooting section ("If your Mk1C does not boot from microSD card") for more information. If you have issues, contact support@nanoporetech.com.

Update to the latest version of MinKNOW for the MinION Mk1C, which supports the latest chemistry, bug fixes and performance enhancements.

Please visit our Oxford Nanopore FAQ and Help Centre if you are having difficulties updating MinKNOW on the MinION Mk1C.

Completing your OS update

All sequencing data stored in the /data folder is retained after the OS upgrade. However, please note that you may need to configure some device settings after the update.

If you use a mounted external storage to the Mk1C, you will need to remount the external storage. Follow the instructions in the Mount network drive section in this user manual.

If you cannot connect to the internet (you may see the error "Unknown connection 'static'"), follow the instructions in the (Advanced) Network configuration - static vs dynamic IP section in this user manual.

20. Updating the signing key to update MinKNOW (IMPORTANT)

IMPORTANT

The MinION Mk1C device has a signing key installed, which is a piece of software used to authenticate MinKNOW updates. On the 16th June 2022 the current signing key has expired, which means you will not be able to update the Mk1C software unless you are either on MinKNOW version 22.05.8, or if you add a new signing key to your device. Below are the instructions for adding the new key to your trusted list via the command line.

To add recognition of the signing key to a MinION Mk1C by command line without also updating the software, connect to the device via SSH.

For details on how to connect via Secure Shell for various operating systems, see the (Advanced) Connecting by the command-line section of this user manual.

Once connected, run the command:

wget -O- https://cdn.oxfordnanoportal.com/apt/ont-repo.pub.2022 | sudo apt-key add -

This command can also be used for MinKNOW for MinION Mk1B Linux installations.

Confirm that you can update MinKNOW via the MinION Mk1C UI after following the steps above.

If you are unable to update MinKNOW or check for updates, repeat the steps above again. If the issue is not resolved, contact Customer Support at support@nanoporetech.com.

21. Export logs

Navigate to the Help section of the Host settings.

Logs

In the 'Export Logs' panel, select 'Export'.

When the logs are successfully exported, the user will be notified of where they are located.

The logs will be downloaded as a TGZ file in the logs directory.

Logs

22. If your Mk1C does not boot from microSD card

Some Mk1C devices with serial numbers MC-110999 or lower may not be able to recognise the microSD card with OS updates. Follow the steps below to change your device so that boot from SD card works.

Copy the file from the link below to your Mk1C.

https://nanoporetech.box.com/s/2vsqb3d6n5mbbtyrbpu1kyai2p3ohknj

Use the following command at your PC/Mac command line/terminal, respectively, to copy the attached file to your Mk1C:

scp ./u-boot.img minit@<your-device>:/home/minit

Note: if you encounter issues, make sure that your PC/Mac is connected via Ethernet to the same network as your MinION Mk1C.

Connect to your device using SSH:

ssh minit@<your-device>

Overwrite the kernel and kernel_b partitions with the image:

sudo dd if=~/u-boot.img of=/dev/disk/by-partlabel/kernel

then run:

sudo dd if=~/u-boot.img of=/dev/disk/by-partlabel/kernel_b

Reboot the device:

sudo reboot now

Leave the device to boot to completion, and do not turn it off mid-reboot.

Last updated: 4/29/2024

Document options

Language: