VolTRAX V2B Technical Specification

Device part number

VOL-V2002b

Device name

VolTRAX V2b

Short description

VolTRAX™ V2b is a small USB-powered device that automates DNA library preparation. VolTRAX is designed to reduce hands-on library preparation time and improve reproducibility of results. Predefined or custom protocols will be made available in future, enabling complete optimisation of sample preparation and the development of novel methods.

Product overview

VolTRAX V2b is an automated, multi-purpose, and portable library preparation device. It improves laboratory efficiency, through advantages over human-made nucleic acid library preparations:

  1. Automation of mixing and separation between reagents;
  2. High levels of reproducibility for nanopore library preparations;
  3. Portability, due to its small size and its lightweight, robust structure;
  4. Providing a controlled environment for incubations and extractions

The above features improve the efficiency of DNA library preparation. Furthermore, it facilitates the biochemical steps required for nanopore sequencing to a new user. Finally, it extends the environmental remit in which a user can prepare a library and/or extract nucleic acid from a biological or environmental sample.

The VolTRAX V2b device

VolTRAX V2b has a metal chassis, containing a heater, peltier, optical fluorescent detector and magnets (the latter four are coloured below). v2 top down coloured features

  • The heater and peltier: (yellow and blue) - These heat the reagents for reactions and incubation stages of the library preparation. The peltier and heater will warm the reagents, while the former can cool the solution temperature too. These enable dual incubations and/or rapid changes between thermal environments.
  • The magnets: (green) - These enable responsive, accurate and contamination-free separation of nucleic acids using magnetic beads.
  • The fluorescent light: (red) - This will enable quantification and quality control checks of nucleic acids. DNA and RNA have a characteristic UV absorption profile at 260 and 280 nm, which is affected by the quality and concentration of the nucleic acids. Appropriate quantities and good quality DNA or RNA are fundamental to a successful sequencing experiment. This feature will directly enable the user to measure the concentration of nucleic acid and assess the quality directly.

The cartridge is supported by a metal chassis, with contact to the four active components shown in the image above. The cartridge has 15 ports and can hold up to 10 different samples for multiplexing. The cartridge is primed by a solution that enables efficient movement and mixing of the loaded libraries and reagents. Cartridge - tech doc

The device also comes with a resin-based lid, which will enable fluorescent detection of nucleic acids.

The samples are loaded into a disposable cartridge that interfaces with a reusable core. On the top surface of the core lies the physical analysis tools for manipulation of the reagents in the cartridge. The tools include a thermal cycler, nucleic acid quantifier (optical fluorescent detector), magnetic bead separator and independent, multi-sample mixing technology.

The cartridge contains an array of individually-addressable pixels. By applying a charge, reagent and sample droplets are moved in a path programmed by software, allowing droplets to be moved, mixed and separated.

The VolTRAX V2b device controls a number of physical/analysis tools such as:

  • A heater and heat pump for thermal incubations and PCR
  • Optical fluorescent light for nucleic acid quantification (this feature will be enabled in future versions of the device software)
  • Magnetic control for nucleic acid-associated magnetic bead separation

Liquid droplets can be held by the carefully configurated cartridges and the VolTRAX V2b electronics to enable the use of these tools for library preparation processes. The VolTRAX V2b device-cartridge interface provides the functions. The VolTRAX V2b interfaces lie on the top face shown below. V2 top with LEDs on - 07 Jan 19 dropbox

Technical specifications

Component Specifications
Size and weight 65 mm x 58 mm x 134 mm, 450 g
Power 15 W
Ports 1x USB Type-C
Pre-loaded software Customer driver
Environmental conditions Designed to prepare a library in environmental temperatures of +18°C to +25°C
Do not cover vents on the top or sides of the device.

Shipping and logistics

The Oxford Nanopore Technologies VolTRAX V2b device is stored and shipped at ambient temperature (15-25°C).

Please note that the VolTRAX V2b is shipped separately to the kits and flow cells.

The delivery charges are calculated when a quote is raised or during checkout. Once an order is made, the delivery ID and delivery information can be tracked in the Store.

IT requirements

VolTRAX V2b is driven by the VolTRAX software. This is responsible for the control of the device and consumable cartridge array.

VolTRAX V2b is less demanding than MinION™. If you are currently running a MinION on a Windows system, or you plan to use a MinION with VolTRAX V2b, the computer requirements for MinION should be all you need to drive VolTRAX V2b.

If you plan to purchase a system dedicated to VolTRAX V2b, the minimum requirements are:

Component Minimum specification
Operating system Windows 10
Mac OS X and Linux are in development.
Memory/RAM 2 GB RAM
CPU 64 bit dual core processor
Storage/hard drive >128 GB SSD
User account privilege level Local administrator
Ports USB TYPE C, 5 V, 3 A †
Antivirus software Installation highly recommended
Internet connection An internet connection and appropriate firewall settings are required for downloading and running the VolTRAX V2b software.

† Only the USB TYPE C – Gen 3.1 (I or II) or higher – 5 V, 3 A can provide the full range of functionalities on VolTRAX V2b.

Safety and legal info

Intended use of the VolTRAX V2b device

Oxford Nanopore Technologies® VolTRAX V2b device is an electronic preparation and analysis system for use in scientific research.

This product is for research use only

The safety information below provides the user with the details needed to install and use the system safely.

Electrical information

Supply voltage 5 VDC, 3 A via USB Type-C – Gen 3.1 (I or II)

Only the USB Type-C – Gen 3.1 (I or II) or higher – 5 V, 3 A can provide the full range of functionalities on VolTRAX V2b
Operating current Up to 3 A (dependent on PCR functionality load)
Maximum power 15 W

Labels on the instrument

VolTRAX V2b device label

VolTRAX V2b carton label

Emergency procedures

In case of emergency, switch the VolTRAX V2b off at the power source by unplugging the USB power cable from the device.

Declaration of conformity

The VolTRAX V2b conforms to the EMC and Electrical Safety directives as outlined in the EC Declaration of Conformity.

2024 DoC VolTRAX V2b

Software license and device warranty

The software licence and device warranty contract ensures your instrument is performing optimally by providing the latest up-to-date hardware and software. The contract guarantees that Oxford Nanopore Technologies support obligations are delivered during the contract period as laid out in sections 4 and 7 of the Nanopore Product Terms and Conditions.

This includes:

  • Software updates upon release
  • Hardware updates on release
  • Return and Replace policy

The service contract extends our warranty to cover the instrument after your initial purchase contract has expired.

What’s in the box

The VolTRAX V2b Starter Pack includes:

  • 1 VolTRAX V2b device
  • 1 Configuration Kit (2 cartridges and blue dye)
  • 4 VolTRAX Cartridge Pack (each pack contains 3 cartridges)
  • 4 VolTRAX Sequencing Kits (each kit contains 3 reactions)
  • 6 MinION Flow Cells
  • 1 USB3.1 / USBC cable

Product cross-compatibility

The VolTRAX V2b can be used together with:

Cartridges

  • Cartridge Pack (VCT-V2002D)

Kits

  • VolTRAX Configuration Kit (VCK-V2003b)
  • VolTRAX Sequencing Kit (VSK-VSK004)
  • VolTRAX PCR Tiling 1-12 COVID-19 Kit (VSK-PTC001)

Change log

Date Version Changes made
24th April 2024 V4 - Corrected values in "Technical specifications"
- Added a Declaration of Conformity
28 July 2022 V3 Updated kit compatibilities
01 December 2021 V2 - Updated the device name to VolTRAX V2b
- Updated the "Short description" to say "VolTRAX™ V2b is a small USB-powered device that automates DNA library preparation."
- Updated the cartridge and kit compatibilities
- USB compatibility statement now says "Only the USB TYPE C – Gen 3.1 (I or II) or higher – 5 V, 3 A can provide the full range of functionalities on VolTRAX V2b"
- Updated the device labels, and removed the image of the label on the SpotON flow cell.
- Environmental conditions have been updated to "Tested to function at 5°C to +40°C."

Last updated: 4/24/2024

Document options